590775
20
Zoom out
Zoom in
Previous page
1/72
Next page
20 GMC-I Messtechnik GmbH
Combined Testing – Differential Protective Conductor Resistance
Zero balancing is also possible for protective conductor measurement. With
zero balancing, all subsequent measurements are adjusted with an offset
such that 0 Ω is displayed for a selected reference point which is con-
nected to the protective conductor. When test points are contacted with
the probe which are electrically connected to this reference point, differen-
tial resistance ΔR
PE
between the reference point and the contacted test
point is displayed. The mains release key must be activated during
measurement in order to perform zero balancing. Press key
„Store
value“ in order to save the reference and/or correction value. The meas-
sage „Zero point corrected“ concerning the reference value is displayed
during all future measurements.
Attention: It is absolutely essential to delete the reference value after the
reference value has been stored and the test has been performed as it is
taken into account in all future tests. To delete the value, the procedure is
the same as for storage, press key
„Delete value“.
9.1 Maximum Allowable Limit Values for Protective Conductor Resistance
for Connector Cables with a Length of up to 5 m
1)
This value may not exceed 1 Ω for permanently connected data processing
systems (DIN VDE 0701-702, DIN VDE 0701, part 240).
2)
Permanently connected cable
3)
Feature G00 = 10 A (SECULIFE ST) / G01 = 25 A (SECULIFE ST HV)
10 Insulation Measurement
10.1 Insulation Resistance R
INS
.
Definition
Safety Class I
Insulation resistance is
measured between short-
circuited mains terminals and
the protective conductor.
Safety Classes II and III
Insulation resistance is
measured between short-
circuited mains terminals and
external conductive parts
which can be contacted with
the probe.
Test Sta ndard Test Current
Open-Circuit
Voltage
R
PE
Housing –
Device Plug
R
PE
Housing –
Mains Plug
VDE 0701-
0702:2008
> 200 mA
4 V < U
L
< 24 V
0.3 Ω
1)
+ 0.1 Ω
2)
additional 7.5 m
IEC 62353
(VDE 0751-1)
EN 61010
10 A
~/25 A
3)
to test socket
only
0.2 Ω 0.3 Ω
0.1 Ω
0.2 Ω
EN 60335
EN 60950
EN 60601 0.2 Ω
2)
Iso / HV
Ω
f
o
r
S
C
I
I
20


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info