GMC-I Messtechnik GmbH 21
Exception for Permanently Installed Safety Class I Devices
Attention!
!
Deactivate the electrical system which supplies power to the
device under test before connecting the test instrument!
õ Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
õ Connect the probe to phase conductor L at the device under test in
order to measure insulation resistance.
Sequence
Attention!
!
Measurement of Insulation Resistance (equivalent leakage current)
This test is performed with a maximum voltage of 500 V, with a
current limit having been set (I < 3.5 mA). However, contacting
the terminals (3 or 2) causes an electric shock which, in turn,
may result in accidents.
Note
+
All switches at the device under test must be set to the ON
position during measurement of insulation resistance. This also
applies to temperature controlled switches and temperature
regulators.
Measurement must be performed in all program stages for
devices equipped with program controllers.
R-INS
Start measurement.
Nominal voltage is 500 V DC in this
case.
Nominal voltage can be adjusted within
a range of 50 V to 550 V DC.
Note
+
When insulation is first started from the menu, nominal voltage is
always set to 500 V. Open-circuit voltage is always greater than
nominal voltage.
Minimum Allowable Limit Values for Insulation Resistance
* for Safety Class I devices with activated heating elements
Notes
All exposed, conductive parts of safety class II and III devices, as well as
of battery-powered devices must be scanned with the probe and insula-
tion resistance and/or leakage current must be measured.
Batteries must be disconnected from their terminals during testing of bat-
tery powered devices.
Test Standard Test Voltage
R
INS
SCI SCII SCIII Heat
VDE 0701-
0702:2008
500 V
1MΩ 2MΩ 0,25 MΩ 0.3 MΩ
*
IEC 62353
(VDE 0751-1)
2MΩ 7MΩ
70 MΩ 70 MΩ
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info