9 Measuring Protective Conductor Resistance
Definition
Protective conductor resistance
is the sum of the following
resistances:
Connector cable or device
connector cable resistance
Contact resistance at plug
and terminal connections
Extension cable resistance if
utilized
Resistance is measured:
Between each conductive
part of the housing and the earthing contacts at the mains and the de-
vice plug (if a removable mains connector cable is used), or the protec-
tive conductor terminal for permanently installed devices.
as 4-pole measurement
Between the earthing contacts at the mains plug and the earthing
contacts at the device plug for device connector cables
Between the earthing contacts at the mains plug and the earthing
contacts at the coupling socket for extension cables
Connecting Safety Class I Devices to the Test Socket
When the DUT is connected, resistance is measured between the protec-
tive conductor terminal at the test socket or at the PE jack and the probe
connection at the DUT (contact with conductive parts of the housing).
õ In order to measure protective conductor resistance, contact a
conductive part of the housing with the probe, which is connected to
the protective conductor.
During measurement, the connector cable must only be moved in as far as
it is accessible during repair, modification or testing.
If a change in resistance occurs during the manual test step of the conti-
nuity test, it must be assumed that the protective conductor is damaged,
or that one of the connector contacts is no longer in flawless condition.
Testing Extension Cables
See test sequence in chapter 15.7 on page 46.
Note
+
“Connection of the DUT: SC I/II” is not displayed when the test is
performed individually, but rather only during the automatic test
sequence.
Selecting Test Current and Polarity
Test current amperage (200 mA DC or 10 A AC (Feature G00 or SECULIFE ST) or
200 mA DC, 25 A AC (Feature G01 or SECULIFE ST
HV)) as well as polarity can
both be changed by pressing the or the key.
Testing with 10 A Test Current (Feature G00 or SECULIFE ST)
or 25 A (Feature G01 or SECULIFE ST
HV)
The test has a maximum duration of 30 s (fixed value) if 10 A or 25 A test
current is used. After this time period has elapsed, the last measured
value is frozen and “data hold, measurement stopped” appears at the dis-
play. If the test instrument becomes excessively warm, testing cannot be
repeated until after a waiting period of 1 minute. When testing with 10 A or
25 A, the last measurement can be repeated if the test results in failure.
PE
Ω
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info