18 GMC-I Messtechnik GmbH
8 Configuring Device Parameters
General device
parameters can be
configured and
saved with the se-
lector switch in the Setup position.
Select a menu and acknowledge.
Select parameter and acknowledge, change setting and acknowledge.
Limit values ...
Settings x / – = function activated / deactivated
Illumination Background illumination for the LCD. One of three
different conditions can be selected* with the up and
down scroll keys: x: continuously on, –: off
numbers 1 through 9: duration in minutes after which
illumination is automatically deactivated.
Test time Duration of a single test (0 255 s)
Reference voltage: Voltage to which leakage current makes reference
(as a rule nominal line voltage)
Earth fault: During the short-circuit test, testing is also performed
to determine whether or not a connection exists
between L/N and PE (short-circuit to exposed con-
ductive part). We assume that a short-circuit to an
exposed conductive part exists in the event of leak-
age current greater than 15 mA from L/N to PE. This
value should be increased for some DUTs (in particu-
lar high-current consumers), because greater leak-
age currents are present.
Mains wait
Line voltage is first applied to the test socket. However,
testing does not begin until after the duration selected in
seconds under “Mains pause” has elapsed.
* The background illumination of test instruments equipped with panels produced
as of 2014 can no longer be deactivated.
Automode x: for fully automatic test sequences the messeges
are mainly suppressed
Test Sequence ...
Settings x / – = function activated / deactivated
Single-fault
If the single-fault condition has been activated, the test
is interrupted as a failure as soon as an error occurs.
Auto Class PSI Test results (passed or failed) for the various selector
switch positions are automatically assigned to the 8
statistics channels.
inc. Service Error Measurement results are compensated by taking
service error into consideration (measuring error).
IT Network Testing in IT systems can be performed by suppress-
ing tests for U
PE-N
. The U
PE-N
test determines
whether or not voltage is present at PE.
(Leakage current measurement results may other-
wise be distorted.)
Acst Sig, Seq
Acoustic signal is generated for: Incorrect connection of the
DUT, error in the electrical supply system, next test step.
Acst Sig, Meas Acoustic signal is generated for: Measured value
fluctuations, test current polarity reversal
Auto meas. point Prerequisite: Feature KD01*. An acoustic signal indi-
cates whether or not the probe is connected to the
protective conductor. The test sequence is run auto-
matically. Rapid signal frequency: probe connected
to PE; slow signal frequency: next measuring point
Direct Print-Out
Prerequisite: Feature KE01*, see chapter 18 on page 59.
Reports Reports which have been saved to memory can be
selected from a list with an ID number and displayed
(see chapter 18 on page 59).
Select template a report template can be selected for print-out from
of 5 different templates.
SECUSTORE Optimize data transfer for the SECUSTORE adapter
connection (in this configuration, no data can be
saved to the (P)SI module). Neither is a test report
signalled to the RS232 interface.
Service – Time and date settings (if a (P)SI module is used,
the same time and date must also be entered to the
(P)SI menu)
– Service functions (password required)
Setup
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info