22 GMC-I Messtechnik GmbH
10.2 Equivalent Leakage Current
General
The equivalent leakage current measurement is a method for measuring
protective conductor current (DIN VDE 0701-0702) and/or device leakage
current (IEC 62353 (VDE 0751)).
A high-impedance power supply is connected between the short-cir-
cuited mains terminals and all exposed metal parts of the housing (which
are connected to one another).
Measurement of Equivalent Leakage Current I
EL
per DIN VDE 0701-0702
Measurement of Equivalent Device Leakage Current I
EDL
(per IEC 62353
(VDE 0751-1))
The patient ports are linked with
each other and are connected to
the same point on the housing.
Equivalent Patient Leakage Current I
EPL
(IEC 62353 (VDE 0751-1))
Test Setup
A high-impedance power supply
is connected between each pa-
tient port and all exposed metal
parts of the housing (which are
connected to one another). The
mains terminals are short-cir-
cuited, and connected to the
same point on the housing.
Measurement
The current which flows over the
insulation at the device under test is measured separately for each appli-
cation part.
Measurement is always performed with an AC power supply with current
limiting. Varying line voltages are taken into account.
Maximum Allowable Limit Values for Equivalent Leakage Current in mA
I
EL
Equivalent leakage current
I
EDL
Equivalent device leakage current
I
EPL
Equivalent patient leakage current
PE Protective conductor
1)
for devices with heating power of 3.5 kW
2)
with and without line voltage applied to the application part
~
DUT
Probe
mA
PE
L
N
Test s o cket
4-5
Test socket
Probe
PAT
DUT
~
mA
L
N
PE
4-5
Test socket
Test socket
A-K
Test Standard I
EL
I
EDL
I
EPL
VDE 0701-
0702:2008
SC I: 3.5
1 mA/kW
1)
SC II: 0.5
IEC 62353
(VDE 0751-1)
SC II 0.2 Type BF 5
2)
SC I (parts within or connected to PE)
1 Type CF 0.05
2)
Permanently connected devices with PE 10
Portable x-ray devices with additional PE 5
Portable x-ray devices without additional PE 2
Devices with mineral insulation 5
mA
PE
PAT
DUT
L
N
~
Probe
Test socket
4-5
A-K
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info