Operational procedures
31
850-164692 / Rev.B
12 In the Calibration window, open the File menu, choose
New to open the Record dialogue and to start a new
calibration.
13 Enter the following data in the Record dialogue box:
a Transducer’s serial number.
b Correct theoretical TS for the reference target.
c Allowed deviation from the TS for the reference target.
(This is a window giving the limits for the system’s
acceptance of single target echoes coming from fish.
The closer to the correct reference target TS the limits
have been set, the more of the unwanted fish echoes
will be rejected. On the other hand, if the echo from the
reference target is too close to one of the limits, the
deviation has to be increased before starting collecting
data. This is because it will always be a certain
variation in TS values when the reference target is
being moved to cover the complete beam.)
d Upper and lower depth limits for the target window. (A
narrow window will have same effect as above,
reducing the possibility of detecting unwanted echoes
from fish.)
e Any comments you may wish to add to the calibration
file
14 Click OK when you have finished entering data.
The calibration program will now begin recording data.
Views
YouwillseetwoviewsintheCalibration window; a Plot view
and an Information view.
A vertical bar is shown on the left side of each view. A blue
colour indicates that the view is active, while gray colour
indicates a passive view. If you wish to print a view, or perform
other operations connected to it, you must make sure that the
desired view is active.
In the Plot view you will see recorded data plotted as blue and
red circles. Blue circles indicate TS values below the current
beam model, while red circles indicate values above the current
beam model.
In the upper part of the Information view you will see various
information associated with recording of the calibration data.
Lines containing this information all begin with a #. Below this
information, recorded values for each new TS detection are
updated continuously during data recording.
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Simrad ER60 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Simrad ER60 in the language / languages: English as an attachment in your email.

The manual is 3,03 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info