19
Alarms and Limits
Alarms and Limits
This chapter describes alarms and limits for the Avant 9600 Digital Pulse Oximeter.
High Priority Alarms
High priority alarms are those that require immediate attention to the patient. They
include SpO
2,
pulse rate, and low perfusion alarms. On the Avant 9600, high priority
alarms are indicated with rapidly blinking red LED displays when alarm limits are
met or exceeded. In addition, the pulse strength bargraph LED may illuminate a red
segment to indicate low perfusion.
High priority alarms are sounded as follows: “beep, beep, beep,” (short pause),
“beep, beep” (10-second pause).
Medium Priority Alarms
Medium priority alarms are those that signal potential problems with the equipment
or other non-life-threatening situations. On the Avant 9600, medium priority alarms
are indicated with slowly blinking amber displays.
Medium priority alarms are illuminated amber on the Main Alarm LED and on the
appropriate indicator(s) or numeric displays, sometimes displaying an error code to
help the user identify the source of the error.
Medium priority alarms are sounded as follows: “beep, beep, beep,” (25-second
pause), “beep, beep, beep.”
Watchdog Alarms
Watchdog alarms are loud, two-tone, steadily beeping signals that indicate a
hardware or software malfunction. When a watchdog alarm is activated, it can be
cleared by shutting down the Avant 9600. If the watchdog alarm cannot be cleared,
remove the battery and contact your distributor or NONIN Customer Support.
Informational Tones
Informational tones communicate important information. They are typically single
“beeps” or a series of three “beeps.” Informational tones include the startup/
initialization tone and the pulse rate tone (which changes in pitch with SpO
2
values).
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Avant 9600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Avant 9600 in the language / languages: English as an attachment in your email.

The manual is 0,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Avant 9600

Nonin Avant 9600 User Manual - German - 47 pages

Nonin Avant 9600 User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info