Alarm Summary
The Avant 9600 detects both patient and equipment alarms. In general, patient
alarms are identified as high priority, while equipment alarms are identified as
medium priority.
Patient Alarms
If patient SpO
2
or pulse readings are equal to or above the upper alarm limit, or if
they are equal to or below the lower alarm limit, the Avant 9600 will signal a high
priority alarm, indicated by red numeric LEDs that flash in sync with the Main
Alarm LED.
Equipment Alarms
High Priority Alarm Description Default Adjustment Options Step Value
SpO
2
Upper Alarm Limit Off Off, 80 to 100 1% SpO2
SpO
2
Lower Alarm Limit 80% Off, 50 to 95 1% SpO2
Pulse Upper Alarm Limit 200 BPM Off, 75 to 275 5 BPM
Pulse Lower Alarm Limit 50 BPM Off, 30 to 110 5 BPM
Low Perfusion Alarm Red segment on Pulse Strength Bargraph
indicates low patient perfusion.
Alarm Description Visual Indicator
Low Battery Alarm Battery LED blinks in sync with Main Alarm LED.
Sensor Alarm Pulse Oximeter Sensor LED blinks in sync with
Main Alarm LED.
Other Equipment Alarms Amber error code appears in main display area.
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Avant 9600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Avant 9600 in the language / languages: English as an attachment in your email.

The manual is 0,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Avant 9600

Nonin Avant 9600 User Manual - German - 47 pages

Nonin Avant 9600 User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info