4 Operation
GB
4.4
Troubleshooting
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse could be
determined.
Check the connections between the cuff and the main unit.
Err 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
Err 5
An average value could not be calculated due to unstable conditions while the
readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6
There were too many errors during the measurement in MAM mode, making
it impossible to obtain a final result.
Read through the checklist for performing reliable measurements and then
repeat the measurement.
HI
The pressure in the cuff is too high (over 300 mmHg) or the pulse is too high
(over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
If no buttons are pressed for approx. 1 minute, the device automatically switch-
es to standby mode. You can also switch to standby mode by pressing the
START button
. The blood pressure meter functions are now switched off.
The time and the selected user memory are continuously shown on the display.
The alarm function remains active.
If you do not want to use the time display and the alarm function, you can also
completely switch off the device by sliding the ON/OFF switch
on the back
of the device into the “ OFF ” position.
4.5
Switching off
the unit
51080_MTX_West_FINAL_akt.qxd:Blutdruckmessgerät 10.09.2009 11:16 Uhr Seite 30
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medisana MTX 51080 bovenarm at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medisana MTX 51080 bovenarm in the language / languages: English, German, Dutch, French, Italian, Portuguese, Swedish, Spanish, Finnish as an attachment in your email.

The manual is 1,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info