599209
39
Zoom out
Zoom in
Previous page
1/182
Next page
29
GB
4 Operation
On the display, you can see
(average) and “ 1 “ for the first
measurement.
2. Next, the sleeve
automatically inflates until the pressure required for a
measurement has been reached. Alternatively, you can create this pressure
manually by pressing and holding the START button
after the device
has automatically reached a pressure of at least 30 mmHg. When you think
that the device has reached the ideal starting pressure for a measurement,
release the START button
. The increasing pressure is displayed numeri-
cally.
3. The measuring process begins automatically when inflation is complete by
releasing the pressure built up in the cuff. The display shows the current
pressure in the cuff.
4. Initially, deflation occurs at a constant speed. The pulse symbol
ap-
pears on the display. As soon as the pulse can be determined, deflation is
synchronised with the beating of the heart until the measurement ends.
The device beeps for each heartbeat during this procedure, if you did not
switch off the beep function beforehand. At the same time, the pulse
symbol
flashes.
5. The first measurement ends when the pressure has been completely dis-
charged from the sleeve during a 15 second countdown. The countdown
symbol
flashes during this process. Beeps sound during the last 5 sec-
onds.
6. Instead of “ 1 “, the display now shows “ 2 “ for the second measurement.
The second measurement proceeds in the same way as the first (see points
2 - 5).
7. Instead of “ 2 “, the display now shows “ 3 “ for the third measurement.
The third measurement proceeds in the same way as the first (see points
2 - 4).
8. Once the 3MAM measurement has been taken, the sleeve suddenly de-
flates and a longer beep sounds.
9. The measured values
,
and
are shown on the display, together
with the WHO blood pressure bar display
. At the same time, a longer
beep sounds.
10. If the unit detects an arrhythmia during the measuring process, the arrhy-
thmia symbol
will flash eight times while the reading is displayed.
You may also hear 8 short beeps; these warn the user of an irregular heart-
beat.
11. If you do not press any buttons for approx. 1 minute, the device automat-
ically switches to standby mode.
NOTE
When the device is used, the errors Err 1, 2, 3, 5, 6 can occur in
the integrated circuit. Err 6 can only occur in 3MAM mode.
In 3MAM mode, errors Err 1, 2, 3, 5 do not cause the
measurement to be cancelled, unless they occur twice in
succession. The error display is always accompanied by 3 short
beeps.
51080_MTX_West_FINAL_akt.qxd:Blutdruckmessgerät 10.09.2009 11:16 Uhr Seite 29
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medisana MTX 51080 bovenarm at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medisana MTX 51080 bovenarm in the language / languages: English, German, Dutch, French, Italian, Portuguese, Swedish, Spanish, Finnish as an attachment in your email.

The manual is 1,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info