8 GMC-I Messtechnik GmbH
5.3 Measurements with the Guard Cable (with Accessory Guard 5000A)
The measurement of very high impedance resistances presupposes
extremely small measurement currents, and can be rendered problematic
by influences such as electromagnetic fields, moisture or surface currents.
For this reason, the measurement set-up must be implemented in an
accurate fashion.
For measurements within a range of 100 G (10 G) ... 1 T a guard
cable must be used in order to prevent the distortion of measurement
results caused by surface currents. The guard rings prevent current at the
surface of the insulation material from flowing from the +measurement
cable to the measurement cable, instead of through the insulation mate-
rial itself.
Ð Insert the guard cable plug into the appropriate jack at the tester.
Ð Plug the crocodile clip onto the guard cable test prod.
Ð Clip the crocodile clip onto the guard ring located between the two
measuring points of the insulation material to be measured.
Ð See Chapter 5.1, page 6 for measurement procedures.
Note
The following materials can be used as guard rings:
aluminum foil, copper foil or metal tubing clamps.
6 Characteristic Values
Insulation Resistance
ShortCircuit Current I
K
1.3 mA
Direct and Alternating Voltage
Protective Devices
1)
with reference to scale length 97.5 mm (100 M range) or 109.8 mm (1 T range)
2)
PTC thermistor cool-down period until start of new measurement:
at least 2 minutes must be observed!
+Measurement Cable
– Meas. Cable
Guard Cable
Conductor
Insulation
Material
Guard Rings
Contact Ring
Measuring
Range
Nominal
Range of Use
Nominal/Test
Voltage
U
N
/ U
T
Nom./Test
Current
I
N
/ I
T
Intrinsic Un-
certainty
1)
Measuring Un-
certainty
100 k...
100 M
100 k...
10 M
100 V
250 V
500 V
1000 V
1mA
2.5%
30%
of measured
value
10 k... 1 T
100 k...
100 G
100/1500 V
250/
2000 V
500/
2500 V
1000/5000 V
1mA/0.7mA
1mA/0.5mA
1mA/0.4mA
1mA/0.1mA
5%
Measuring Range Frequency Internal
Resistance
Max. Allowable
Voltage
Intrinsic
Uncertainty
1)
0 ... 2000 V DC/AC 15 ... 500 Hz 5 M
2200 V DC/AC
max. 10 s
5%
Terminal Internal
Resistance
Max. Allowable
Voltage
Protective Device
–Measurement
Cable
to +meas. cable/
to guard cable:
2000 V DC/AC
max. 10 s
via grounded
damping diodes
+Measurement
Cable
Insulation
Measurement
to –meas. cable/
to guard cable:
2000 V DC/AC
max. 10 s
diodes in high-voltage
cascade, PTC thermistor
2)
and series resistors
Guard Cable
between guard and
meas. cables 90 k
to meas. cable
2000 V DC/AC
max. 10 s
PTC thermistor
2)
and
series resistors
Battery 10 V
pole protection
with diodes voltage limiting
in battery charger (optional)
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME

Gossen Metrawatt METRISO PRIME User Manual - German - 14 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info