590654
9
Zoom out
Zoom in
Previous page
1/14
Next page
GMC-I Messtechnik GmbH 9
Display
Movement core-magnet moving-coil mechanism
Scale Length 111.5 mm (longest scale)
Reference Conditions
Ambient Temp. +23 C 2K
Relative Humidity 40 ... 60%
Meas. Quantity
Frequency 50 Hz 10 Hz (for voltage measurement)
Mains Voltage
Waveform sine, deviation between effective and
rectified value < 1%
Battery Voltage 8 V1%
Operational Position horizontal
Power Supply
Battery or Storage
Battery 6 ea. 1.5 V single cell per IEC R20 (6 · D-Size)
Working Range 6 V ... 10 V
Battery Service Life 7500 measurements at 1000 V test voltage
with 1 M measuring shunt,
15,000 measurements at 500 V test voltage
with 500 kmeasuring shunt,
with 5 s measurement and 25 s pause
Ambient Conditions
Operating Temp. 0 C ... + 40 C
Storage Temp. 20 C ... + 60 C (without batteries)
Relative Humidity max. 75%, no accumulation of moisture
Elevation to 2000 m
Electrical Safety
Protection Class II
Test Voltage 8.5 kV~
Measurement Category 1000 V CAT II, 600 V CAT III, 300 V CAT IV
Nominal Voltage U
N
1000 V
Open-Circuit Voltage U
0
5000 V
Pullution Degree 2
Electromagnetic Compatibility EMC
Product Standard DIN EN 61326-1: 2006
Mechanical Design
Dimensions W x D x H: 290 mm x 250 mm x 140 mm
Weight 3.4 kg with batteries
Protection IP 52
Extract from table on the meaning of IP codes
Making Capacity for Insulation Resistance Measurement
Response Time < 100 G < 3 s;
> 100 G < 8 s also valid
for test voltage or measuring range changes
Interference emission Class
EN 55022 B
Interference immunity Test value Performance feature
EN 61000-4-2 Contact/atmosphere - 4 kV/8 kV B
EN 61000-4-3 10 V/m B
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
0 not protected 0 not protected
1 50.0 mm dia. 1 vertically falling drops
2 12.5 mm dia. 2
vertically falling drops
with enclosure tilted 15
3 2.5 mm dia. 3 spraying water
4 1.0 mm dia. 4 splashing water
5 dust protected 5 water jets
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME

Gossen Metrawatt METRISO PRIME User Manual - German - 14 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info