590611
25
Zoom out
Zoom in
Previous page
1/28
Next page
GMC-I Messtechnik GmbH 25
Service Life with alkaline-manganese batteries: approx. 100 hours
Battery Test Battery capacity display with battery symbol in
4 segments: “
Battery Saving Circuit
The instrument is switched off automatically
if the measured value remains unchanged for a period of approximately 10
minutes, and if none of the control elements are operated during this period.
Automatic shutdown can be disabled.
If battery voltage drops to below approx. 1.8 V
Fuses
300
A to 10 A range
– Resettable 15A/240VAC/50VDC miniature circuit breaker,
– A fuse link is additionally connected in series to the
auto-fuse, the blowing or absence of which is detected
automatically (“FUSE” is displayed):
T16A/500VAC, 6,3 mm x 32 mm
Switching capacity 1.5 kA at 500 V AC and ohmic load
Electrical Safety
Safety Class II per IEC/EN 61010-1:2010/VDE 0411-1:2011
Measurement Category
CAT I I
Operating Voltage 300 V
Pollution degree 2
Test Voltage 2.3 kV~ per IEC/EN 61010-1:2010/VDE 0411-1:2011
Electromagnetic Compatibility (EMC)
Interference Emission EN 61326-1:2006 Class B
Interference Immunity EN 61326-1:2006
EN 61326-2-1:2006
Ambient Conditions
Accuracy range 0 C
+40 C
Operating Temp. 10 C
+50 C
Storage Temperature 25 C
+70 C (without batteries)
Relative Humidity max. 75%, no condensation allowed
Elevation to 2000 m
Deployment indoors, outdoors: only within the specified
ambient conditions
Mechanical Design
Protection housing: IP 40, connector jacks: IP 20
Table Excerpt Regarding Significance of IP Codes
Dimensions 146 mm x 118 mm x 44 mm
Weight approx. 450 gr. with batteries
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
2 12.5 mm dia. 0 not protected
4 1.0 mm dia. 0 not protected
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAport 40S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAport 40S in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info