13 Temperature Measurement with Pt100 and Pt1000
Set the rotary switch to
“Temp RTD”.
Connect the P
t sensor to
the and V jacks. The
instrument automatically
recognizes type of
connected sensor (Pt100
or Pt1000), and displays
measured temperature in
the selected unit of
measure.
Note!
The characteristic cable resistance for the temperature sensors which
are available as accessories is automatically taken into consideration for
this measurement.
Temperature Measurement
with Compensation for Sensor Cable Resistances from 0.1 to 50
Sensor cable resistances with values other than100 m can be compensated
for values of up to 50 as follows:
Briefly press the FUNC and the MAN/AUTO keys simultaneously. The currently
selected cable resistance is displayed. This value can be increased with the
DATA key, or reduced with the MAN/AUTO key. Each time one of these keys is
briefly activated, the value is changed by 10 digits (0.1 ). The value can be
changed by means of rapid scrolling by pressing and holding the key.
Switch back to temperature measurement by briefly acknowledging with the
FUNC key.
The new cable resistance value is retained, even after the instrument has been
switched off.
Note!
The default setting is Pt100
/Pt1000
with a cable resistance of 0.1 .
14 Temperature Measurement with Type K Thermocouple
Set the rotary switch to “Temp RTD”.
Select “Temp TC K” measurement with thermocouple by repeatedly press-
ing the FUNC key until C or F appears at the display.
Select temperature unit of measure °C or °F by long pressing the FUNC key.
Connect the sensor to the “” and “V” sockets.
Z3409
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAport 40S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAport 40S in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info