16 GMC-I Messtechnik GmbH
8 Resistance Measurement
Set the rotary switch to the
position. Overload is indicated
if no device under test has been
connected: “0.L M”.
Before connecting the device under
test, make sure that it is voltage-
free. Interference voltages distort
measurement results! Perform a
voltage test first if required.
Connect the device under test as
shown in the diagram.
Zero Balancing in the 30 , 300 and 3 kMeasuring Ranges
Resistance at cables, as well as contact resistances, can be eliminated for the
measurement of low-resistance values in the 30 , 300 and 3 k ranges by
means of zero balancing:
Connect the measurement cables to the instrument and connect the free
ends to one another (short circuit the test probes).
Briefly press the FUNC key.
The instrument acknowledges zero balancing with an acoustic signal, and
00
.00 ”, 000.00 ” or “0.0000 kand the ZERO symbol appear at the
LCD.
Resistance measured at the moment the key is pressed is used as a
reference value (max. 2000 digits). This value is automatically subtracted
from all subsequently measured values.
Zero balancing can be cleared:
by pressing and holding the FUNC key, after which clearing is acknowledged
by a twice repeated acoustic signal,
by switching the instrument off.
9 Continuity Testing
With the “acoustic signal” function activated, and exclusively in the 0 to 310
measuring range, the instrument generates a continuous acoustic tone for
measured resistance within a range of 0 to approximately 2 .
Turn the selector switch to the position. The and symbols appear at
the LCD.
Connect the measurement cables to the device under test.
Note!
Continuity testing is very fast (< 50 ms) and is suitable for locating
connections with poor contact (e.g. due to vibration) in automotive
service applications.
R
x
Voltage Drop
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAport 40S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAport 40S in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info