7.4.1 µA DC current range (BENNING MM 6-1)
The input resistance is approx. 3 kΩ.
Overload protection: 1000 V
AC/DC
Measuring range OL indication Resolution Measurement accuracy
600.0 µA 660.0 µA 0.1 µA ± (1.0 % of reading + 5 digits)
7.5 AC current ranges (BENNING MM 6-2)
Overload protection:
- 11 A (1000 V AC/ DC) fuse, 20 kA, rapid on 10 A input
Max. measuring time:
- 3 minutes at > 5 A (pause > 20 minutes)
- 30 secondes at > 10 A (pause > 10 minutes)
Measuring
range
OL indication Resolution
Measurement accuracy
in frequency range 45 Hz - 500 Hz (sine)
6.000 A 6.600 A 0.001 A ± (1.5 % of reading + 5 digits)*
10.00 A 20.00 A 0.01 A ± (1.5 % of reading + 5 digits)*
* 6 A measuring range from 20 mA, 10 A measuring range from ≥ 100 mA
7.5.1 μA AC current range (BENNING MM 6-1)
The input resistance is approx. 3 kΩ.
Overload protection: 1000 V
AC/DC
Measuring
range
OL indication Resolution
Measurement accuracy
in frequency range 45 Hz - 500 Hz (sine)
600.0 µA 660.0 µA 0.1 µA ± (1.5 % of reading + 5 digits)*
* Measuring range from ≥ 1 µA
7.6 Resistance ranges
Overload protection: 1000 V
AC/ DC
Measuring range OL indication Resolution Measurement accuracy
600.0 Ω 660.0 Ω 0.1 Ω ± (0.9 %
of reading + 8
digits
)
6.000 kΩ 6.600 0.001 ± (0.9 %
of reading + 5
digits
)
60.00 kΩ 66.00 0.01 kΩ ± (0.9 %
of reading + 5
digits
)
600.0 kΩ 660.0 0.1 kΩ ± (0.9 %
of reading + 5
digits
)
6.000 MΩ 6.000 0.001 ± (0.9 %
of reading + 5
digits
)
40.00 MΩ 40.00 MΩ 0.01 MΩ ± (1.5 %
of reading + 8
digits
)
*
* Measured values > 10 might cause changing values on the display
(max. ± 50 digits).
7.7 Diode testing
Overload protection: 1000 V
AC/ DC
Max. no-load voltage: 1.8 V
Measuring range OL indication Resolution Measurement accuracy
1.500 V 1.550 V 0.001 V ± (0.9 %
of reading + 5
digits
)
7.8 Continuity testing
Overload protection: 1000 V
AC/ DC
The built-in buzzer sounds in the case of a resistance R less than 20 Ω up to 200
Ω. For a resistance R higher than 200 Ω, the buzzer does not emit an acoustic
signal. Additionally, in case of continuity, the red LED
N
on the top side of the
device lights.
Measuring range OL indication Resolution Measurement accuracy
600.0 Ω 660.0 Ω 0.1 Ω ± (0.9 % of reading + 8 digits)
7.9 Capacity ranges
Conditions: capacitors discharged and connected in accordance with the polar-
ity stated.
Overload protection: 1000 V
AC/ DC
Measuring range OL indication Resolution Measurement accuracy
1.000 µF 1.100 µF 0.001 µF ± (1.9 % of reading + 8 digits)
10.00 µF 11.00 µF 0.01 µF ± (1.9 % of reading + 5 digits)
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning MM6.2 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning MM6.2 in the language / languages: English, German, Dutch, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 2,95 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info