IM 1F5B2-01E
8. MAINTENANCE
8.3 Disassembly and Reassembly
When disassembling or reassembling ULTRA YEWFLO for a part replacement or
maintenance, ALWAYS turn off both the flowmeter and flow of the process fluid. Use
the appropriate tools for the disassembly/reassembly of ULTRA YEWFLO.
CAUTION
The maintenance work described hereinafter can be done at the location where
ULTRA YEWFLO is installed if it is a standard model. If your ULTRA YEWFLO is
an explosion-protected model, however, it is an established rule that you relocate it to
a non-hazardous area, do the maintenance work, and then reassemble it in its original
form. See “Installation and Operating Precautions for Flameproof Explosion-Pro-
tected Instruments” at the end of this manual for more information.
No user modification is permitted on flameproof models of ULTRA YEWFLO. You
therefore are not allowed to add any additional indicator/totalizer component to those
models of ULTRA YEWFLO nor can you use ULTRA YEWFLO with the compo-
nent removed. If such modification is unavoidably necessary, consult Yokogawa
Electric Corporation.
When disassembling a flameproof model of ULTRA YEWFLO, first turn the lock
screw clockwise with a special tool (Allen wrench) to unlock the cover, and then
remove the cover. After having replaced the cover, ALWAYS turn the lock screw
counterclockwise to lock the cover.
8.3.1 Removing the Indicator/Totalizer Component
When removing the indicator/totalizer component to change the way it is viewed,
follow the procedure described below.
<1> Turn off ULTRA YEWFLO.
<2> Remove the cover on the converter.
<3> Using a flat-blade screwdriver, remove the two setscrews from the indicator/
totalizer component.
<4> Take the component out by hand.
<5> When reassembling the component into the converter after the component is
repositioned, follow these steps in their reverse order, while aligning the
component’s screw holes with the bolts in the converter and then fastening the two
set screws.
57


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Yokogawa UYF200 Ultrasonic Vortex Flowmeter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Yokogawa UYF200 Ultrasonic Vortex Flowmeter in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info