Possible cause Corrective action and explanation
quickly. The American Heart Association recommends
deflation no faster than 3 mmHg per second.
Only use a sphygmomanometer that is calibrated.
An uncalibrated sphygmomanometer may result in
inaccurate blood pressure measurements.
Poor auscultatory sound recognition by observer Use high-quality stethoscope. Have a different observer
check patient’s blood pressure.
Cuff inflation and deflation with no blood pressure reading
displayed
Possible cause Corrective action and explanation
Leak in pneumatic system
Inflation too quick; check NIBP cuff and tubing
connections may display.
Cuff pressure limits exceeded. Powering down may
display.
NIBP air leak; check cuff and tubing connections may
display.
Unable to determine NIBP; check inflation settings
may display.
Ensure all cuff attachments are tight.
Carefully check for leaks in the blood pressure cuff, tubing,
and pressure hose attached to ProBP 3400.
Patient's measured blood pressure is outside the rated
specification limits of the device
Determined blood pressure is outside of the rated range.
Evaluate the patient using manual methods.
Arm movement during cycle
Unable to determine NIBP; check connections; limit
patient movement may display.
Keep arm still during blood pressure cycle.
Movement may cause inaccuracies from artifact, long cycle
times, and error message.
Cuff tubing or pressure hose movement artifact
Unable to determine NIBP; check connections and
tubing for kinks may display.
Do not contact cuff tubing or pressure hose during blood
pressure cycle.
Movement may cause inaccuracies from artifact.
User may have pressed the Blood Pressure Start/
Stop button
User cancelled NIBP reading will display. Press the
Blood Pressure Start/Stop button to initiate
blood pressure measurement.
64
Troubleshooting Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
70


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn ProBP 3400 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn ProBP 3400 in the language / languages: English as an attachment in your email.

The manual is 0,95 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn ProBP 3400

Welch Allyn ProBP 3400 User Manual - German - 82 pages

Welch Allyn ProBP 3400 User Manual - Dutch - 82 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info