Troubleshooting
Inaccurate blood pressure readings
Possible cause Corrective action and explanation
Incorrect cuff size
Use Welch Allyn approved cuffs only.
Measure patient’s arm circumference midway between
elbow and shoulder (see “Blood pressure cuff selection” to
select correct cuff size).
Patient’s arm position Ensure patient’s arm is at heart level.
Arm movement during blood pressure cycle
Keep arm still during blood pressure cycle.
Movement may cause inaccuracies from artifact.
Blood pressure taken over clothing Take blood pressure on a bare arm.
Arrhythmia
Check for regularity of heart rate (palpate pulse or check
device).
Moderate to severe heart rate irregularities may make
blood pressure difficult to measure accurately.
Change in blood pressure between auscultatory reading
and ProBP 3400 reading
Check blood pressure immediately before ProBP 3400
reading.
Blood pressure is dynamic and changing. It is normal for
blood pressure to fluctuate 5 to 10 mmHg.
Incorrect reference Use the correct Korotkoff sound to determine diastolic
blood pressure.
Many listeners incorrectly equate diastolic blood
pressure with the disappearance of sound only (phase
5). ProBP 3400 was developed using the American
Heart Association recommendations, which state that
phase 5 be used unless sound continues to 0 mmHg,
in which case the change in the quality of sound
(phase 4) is to be used.
Deflate cuff no faster than 3 mmHg per second.
One of the major sources of error in auscultatory
blood pressure measurement is deflating the cuff too
63
69


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn ProBP 3400 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn ProBP 3400 in the language / languages: English as an attachment in your email.

The manual is 0,95 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn ProBP 3400

Welch Allyn ProBP 3400 User Manual - German - 82 pages

Welch Allyn ProBP 3400 User Manual - Dutch - 82 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info