622288
10
Zoom out
Zoom in
Previous page
1/104
Next page
DVM1500
V. 03 04/11/2013 10 ©Velleman nv
continuously and the resistance value is displayed. If the measured
resistance exceeds the current range or in case of an open circuit, the
OL
Note: Never perform continuity measurements on a live circuit and make
sure all capacitors are completely depleted.
Diode Measurements
Do not perform diode measurements on live circuits.
1.  °CHz

2. Set the rotary switch to .
3. Press the FUNC button to select diode test .
4. Connect the red test probe to the anode of the diode and the black test
probe to the cathode.
The meter displays the forward voltage drop of the diode. If the diode is
connected with a reversed polarity
Notes
Never perform diode measurements on a live circuit and make sure all
capacitors are completely depleted.
Measuring diodes that are part of a circuit might produce faulty results.
Consider unmounting from the circuit.
Capacitance Measurements
Do not perform capacitance measurements on live circuits.
1.  °CHz

2. Set the rotary switch to .
3. Press the RANGE button to manually select the range.
4. Connect the test probes to the capacitor and read its value from the
display.
Notes:
Depending on the capacitance, the meter needs 5 to 30 seconds to show
the result. This is normal behaviour.
To increase accuracy when measuring low capacitance values (< 4 nF), first
hold the tips of the test probes together to determine the capacitance value
of the test leads. Subtract this value from the measured value of the circuit.
Never perform capacitance measurements on a live circuit and make sure
all capacitors are completely depleted.
Frequency Measurements
Do not perform frequency measurements in circuits with
voltages > 250 V.
10


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Velleman DVM1500 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Velleman DVM1500 in the language / languages: English, German, Dutch, French, Polish, Portuguese, Spanish as an attachment in your email.

The manual is 1,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info