All the alarm levels, including physiologic alarms and technical alarms, cannot be changed by users.
6.3 Alarm Indicators
When an alarm occurs, the monitor will indicate it through the following means:
Alarm tone: According to the alarm level, alarm sounds in different tones will emit from the speaker.
Alarm Light: According to the alarm level, the alarm LED light on the monitor will flash in a different color and speed.
Alarm message: Alarm messages will be displayed on the screen.
Caution: The exact nature of the alarm depends on the specific alarm level.
6.3.1 Alarm Tones
The device will make the following sounds for different level alarms:
Alarm level
Audible prompt
High
“DO-DO-DO------DO-DO, DO-DO-DO------DO-DO”
Medium
“DO-DO-DO”
Low
“DO-
6.3.2 Alarm Lamp
The device has two alarm lamps. One flashes as red/yellow and the other flashes as cyan. When a physiologic alarm
occurs, the alarm levels are indicated in the following visual ways:
Alarm level
Visual prompt
High
Alarm LED flashes red at 2 Hz intervals.
Medium
Alarm LED flashes yellow with 0.5 Hz intervals.
Low
Alarm LED lights up yellow without flashing.
When a technical alarm occurs, the alarm levels are indicated in following visual ways:
Alarm level
Visual prompt
High
Alarm LED flashes red at 2 Hz intervals.
Medium
Alarm LED flashes yellow with 0.5 Hz intervals.
Low
Alarm LED lights up cyan without flashing.
Caution: When multiple alarms of different levels occur at the same time, the monitor will issue visual and
audible alarm indicators for the highest-level issues. When the low level technical and physiologic alarms
occur at the same time, the two alarm lamps will be illuminated simultaneously: one flashes yellow and one is
cyan.
Upper alarm limit exceeded
Low
Lower alarm limit exceeded
Low
TEMP Module Failure
Low
48


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for SunTech CT 50 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of SunTech CT 50 in the language / languages: English as an attachment in your email.

The manual is 2,88 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of SunTech CT 50

SunTech CT 50 User Manual - German - 85 pages

SunTech CT 50 User Manual - French - 86 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info