SECTION III OPERATION MANUAL
Utrecht
- 32 - Ref 002
FLICK Flicker Analysis
Use the SELECT key to enable and/or disable analysis.
ON Enabled Analysis
OFF Disabled Analysis
PER. Periodic Analysis
Use the SELECT key to enable and/or disable analysis.
ON Enabled Analysis
OFF Disabled Analysis
ANOM. WINDOW
Anomaly window
Toggling between OFF, FIXED or VARIABLE recording is possible in
PERIODICS (see Section III, paragraph 3.2.1) or using a PC software.
Voltage Anomaly recording is available only for the voltages
selected for recording (see 3.2.4 SIGNALS) regardless of the status
of EN 50160 Analysis. If no voltage is selected, there will be no
logging of Voltage Anomalies.
OFF Disabled anomaly window recording.
FIXED The window (and the Upper & Lower Limits) is set around the nominal
voltage and remains fixed during recording session.
VARIABLE
The window (and the Upper & Lower Limits) is set around the average
of dynamically calculated voltage. Use the LEFT and RIGHT keys to
adjust the averaging period for calculating new values of average
voltage (1 s to 900 s).
MAIN INTEG.
PER.
Main Integration period
Selected duration for Periodic Analysis.
Use the LEFT and RIGHT keys to set the integration period (between 1
s and 30 min).
POWER SUB. I.P. Power sub Integration period
This function is not active in EN 50160 Recording mode
NOMINAL
VOLTAGE
The nominal voltage used as a reference in Voltage Anomaly recording.
In FIXED window mode, this is the actual voltage used.
In VARIABLE window mode, this is the start value of voltage, later
modified to the average value of voltage during the previous Integration
Period while recording.
This value can be changed in METER Configuration Menu only.
UPPER LIMIT
LOWER LIMIT
These are the limits that define the pass window for Voltage Anomaly
recording. Any voltage value outside the specified limits is detected and
stored as an anomaly.
Use the LEFT and RIGHT keys to set the required limit:
- 1 % to 30 % of nominal voltage for upper limit and
- - 1 % to - 30 % of nominal voltage for lower one.
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nieaf-Smitt PQA 6600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nieaf-Smitt PQA 6600 in the language / languages: English as an attachment in your email.

The manual is 1.72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info