MI 3360 OmegaPAT XA
Bijlage E
72
OUT_1
Instellen NC relais contact tussen OUTPUTS pins 4 en 9
OUT_2
Instellen NC relais contact tussen OUTPUTS pins 3 en 8
OUT_3
Instellen NC relais contact tussen OUTPUTS pins 2 en 7
OUT_4
Instellen NC relais contact tussen OUTPUTS pins 1 en 6
OUT_5
OUT_6
OUT_7
OUT_8
Alleen beschikbaar in combinatie met CE Adapter A 1460
12.4.6.3 Wacht op ingang mode
Leest ingangssituatie op pins IN_2, IN_3, IN_4 en IN_5 on INPUTS poorten. Ingang moet “hoog” zijn om
verder te gaan met Auto Sequence.
Parameters
Status
Aan vrijgeven wacht op ingang mode; Selecteer actieve ingangen in Input pins
menu.
Uit blokkeren Wacht op ingang menu.
IN_2
IN_2 leest of conditie op INPUTS pin 6 actief is
IN_3
IN_3 leest of conditie op INPUTS pin 7 actief is
IN_4
IN_4 leest of conditie op INPUTS pin 8 actief is
IN_5
IN_5 leest of conditie op INPUTS pin 4 actief is
12.4.6.4 Lampen HS mode
Stuurt externe lampen via OUT_1 en OUT_2 uitgangen aan. Alleen actief bij HS en HS- programmeerbaar
functies.
Rode lamp (OUT_1) Aan betekent dat instrument gereed staat voor HS test. Rode lamp gaat aan vóór
eerste flow commando in meetstap met HS-test. Rode lamp gaat uit na einde van HS-test.
Groene lamp (OUT_2) Aan knipperend, betekent dat hoge spanning zal worden doorgegeven aan HS
test aansluitklemmen, zodra volledig voldaan is aan alle ingangs voorwaarden.
Groene lamp (OUT_2) Aan, betekent dat gevaarlijke testspanning op de HS test aansluitingen staat.
Groene lamp gaat voor de meting aan en schakelt uit na de meting.
Als Lampen HS mode commando is vrijgegeven, worden de instellingen voor het Drive uitgangscommando
voor OUT_1 en OUT_2 genegeerd.
Parameters
Status
Aan Vrijgeven Lampen HS mode
Uit Blokkeren Lampen HS mode
12.4.6.5 Lampen Goed/Fout mode
Stuurt externe lampen aan via OUT_3 en OUT_4 uitgangen.
Tijdens de meting geven de lampen de status weer van de enkelvoudige test.
72


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3360 - OmegaPAT XA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3360 - OmegaPAT XA in the language / languages: Dutch as an attachment in your email.

The manual is 2.67 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3360 - OmegaPAT XA

Metrel MI 3360 - OmegaPAT XA User Manual - English - 154 pages

Metrel MI 3360 - OmegaPAT XA User Manual - German - 172 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info