MI 3121H Smartec 2,5 kV Insulation / Continuity Insulation resistance
24
Figure 5.5: DAR result displayed during
measurement (result flashing)
Figure 5.6: PI result displayed during
measurement (result flashing)
After measurement is finished it is possible to switch between DAR, PI, also R60 and
insulation results. R60 is the resistance value measured 1minute (60 seconds) after start
of the measurement.
Simply keep pressing (few seconds) the TAB key to enter to the sub results menu. Here
you can list sub results (DAR, PI, and R60) with normal pressing on the TAB key.
Automatically turn back from sub result menu is occurred, after all sub results are
displayed once. See the Figure 5.7, Figure 5.8 and Figure 5.9.
Figure 5.7: DAR result
displayed after measurement
Figure 5.8: PI result
displayed after measurement
Figure 5.9: R60 value
displayed after measurement
For additional information regarding PI and DAR diagnostic, please refer to Metrel’s
handbook Modern Insulation Testing.
5.1.3 Guard terminal
The purpose of the GUARD terminal is to lead away potential leakage currents (e.g.
surface currents), which are not a result of the measured insulation material itself but of
the surface contamination and moisture. This current interferes with the measurement i.e.
the Insulation Resistance result is influenced by this current. The GUARD terminal is
internally connected to the same potential as the negative test terminal (black one). The
GUARD test alligator should be connected to measured object so as to collect most of the
unwanted leakage current, see the figure below.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3121H SMARTEC 2.5 kV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3121H SMARTEC 2.5 kV in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3121H SMARTEC 2.5 kV

Metrel MI 3121H SMARTEC 2.5 kV User Manual - German - 44 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info