MI 3121H Smartec 2,5 kV Insulation / Continuity Insulation resistance
22
Test circuits for insulation resistance
L1
L2
L3
N
PE
mains voltage
switched off
closed
switches
loads disconnected
Figure 5.2: Connection of 2-wire test lead
L1
L2
L3
N
PE
mains voltage
switched off
closed
switches
loads disconnected
Figure 5.3: Connection of optional 3-wire test lead (A 1319)
Insulation resistance measuring procedure
Select the INS function using the BACK / NEXT keys of function selector.
Set the required test voltage (Press the TAB key then use arrows).
Enable and set limit value (optional).
Disconnect tested installation from mains supply (and discharge insulation as
required).
Connect test lead to the top of the instrument and to the item to be tested (see
Figure 5.2 and Figure 5.3).
Press the TEST key to perform the measurement (double click for continuous
measurement and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged.
Store the result by pressing the MEM key (optional).
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3121H SMARTEC 2.5 kV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3121H SMARTEC 2.5 kV in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3121H SMARTEC 2.5 kV

Metrel MI 3121H SMARTEC 2.5 kV User Manual - German - 44 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info