MI 3121H Smartec 2,5 kV Insulation / Continuity Insulation resistance
25
I
L
I
L
I
L
Ut Ut
I
A
I
A
I
M
I
M
A
A
+OUT +OUT
-OUT
-OUT GUA RD
Figure 5.10: Connection of GUARD terminal to measured object
where:
Ut........Test voltage
IL .........Leakage current (resulted by surface dirt and moisture)
IM ........Material current (resulted by material conditions)
IA......... A-meter current
Result without using GUARD terminal: RINS = Ut / IA = Ut / (IM + IL) …incorrect result.
Result using GUARD terminal: RINS = Ut / IA = Ut / IM ……correct result.
It is recommended to use the GUARD connection when high insulation resistance
(>10G Ω) is measured. You need optional 2.5 kV 3-wire test lead (Figure 5.3).
The guard terminal is internal impedance (660 kΩ) protected.
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3121H SMARTEC 2.5 kV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3121H SMARTEC 2.5 kV in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3121H SMARTEC 2.5 kV

Metrel MI 3121H SMARTEC 2.5 kV User Manual - German - 44 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info