MI3002 MI3100 MI3102 Reduced low voltage supply systems
102
C.4.2 Fault loop resistance and prospective fault current
R LOOP (Rs in the UK version) sub-function
Measuring range according to EN61557-3 is 0.32 Ω ÷ 1999 Ω.
Measuring range (Ω) Resolution (Ω) Accuracy*)
0.00 ÷ 19.99 0.01
20.0 ÷ 99.9 0.1
100 ÷ 1999 1
±(10 % of reading + 5 digits)
*)The accuracy is valid if mains voltage is stable during the measurement.
Prospective fault current (calculated value)
Measuring range (A) Resolution (A) Accuracy
0.00 ÷ 19.99 0.01
20.0 ÷ 99.9 0.1
100 ÷ 999 1
1.00k ÷ 9.99k 10
10.0 ÷ 24.4k 100
Consider accuracy of fault
loop resistance
measurement
IPFC calculation................................ IPFC=UN × kSC / ZL-PE
U
N = 55 V, (44 V U < 61 V) for selected 55 V single-phase system
U
N = 63 V, (56 V U < 70 V) for selected 63 V three-phase system
Test current ...................................... 1.9 A (10 ms)
Nominal input voltage ....................... 55 V / 63 V (45 Hz ÷ 65 Hz)
Test posibilities ................................. L1-PE and L2-PE
Rs(rcd) trip-lock sub-function
Measuring range according to EN61557 is 0.85 Ω ÷ 1999 Ω.
Measuring range (Ω) Resolution (Ω) Accuracy *)
0.00 ÷ 19.99 0.01 ±(10 % of reading + 15 digits)
20.0 ÷ 99.9 0.1 ±10 % of reading
100 ÷ 1999 1 ±10 % of reading
*) Acccuracy may be impaired in case of heavy noise on mains voltage.
Prospective fault current (calculated value)
Measuring range (A) Resolution (A) Accuracy
0.00 ÷ 19.99 0.01
20.0 ÷ 99.9 0.1
100 ÷ 999 1
1.00k ÷ 9.99k 10
10.0 ÷ 24.4k 100
Consider accuracy of fault
loop resistance
measurement
IPFC calculation................................ IPFC=UN × kSC / ZL-PE
U
N = 55 V, (44 V U < 61 V) for selected 55 V single-phase system
U
N = 63 V, (56 V U < 70 V) for selected 63 V three-phase system
Nominal input voltage ....................... 55 V / 63 V (45 Hz ÷ 65 Hz)
Test posibilities ................................. L1-PE and L2-PE
No trip out of RCD.
102


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3102 BT - Eurotest XE at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3102 BT - Eurotest XE in the language / languages: English as an attachment in your email.

The manual is 1.18 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3102 BT - Eurotest XE

Metrel MI 3102 BT - Eurotest XE User Manual - Dutch - 33 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info