MI3002 MI3100 MI3102 Reduced low voltage supply systems
100
C.4 Technical specifications
Only those technical specifications are listed below that are different to specifications
from chapter 8 of this document.
C.4.1 RCD testing
General data
Nominal residual current................... 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy.... -0 / +0.1IΔ; IΔ = IΔN, 2×IΔN, 5×IΔN
-0.1IΔ / +0; IΔ = ½×IΔN
Maximum nominal differential
current for declared accuracy ........... 1000 mA for IΔN
500 mA for 2×IΔN
100 mA for 5×IΔN
Maximum test current ....................... 1 A (for Z LOOP < 1 Ω)
Test current shape............................ Sine-wave (AC), pulsed (A)
DC offset for pulsed test current ....... 6 mA typical
RCD type .......................................... general (G, non-delayed), selective (S, time-
delayed)
Test current starting polarity ............. 00 or 1800
Nominal input voltage ....................... 55 V / 63 V (45 Hz ÷ 65 Hz)
Contact voltage
Measuring range according to EN61557-6 is 3.0 V ÷ 32.6 V for limit contact voltage 25
V.
Measuring range according to EN61557-6 is 3.0 V ÷ 66.0 V for limit contact voltage 50
V. Measuring range (V) Resolution (V) Accuracy
0.0 ÷ 9.9 0.1 (-0 % / +15 %) of reading +
2 digits
10.0 ÷ 99.9 0.1 (-0 % / +15 %) of reading
Accuracy applies for 1 year in reference conditions. Temperature coeficient ouside
these limits is +1 digit.
Test current ...................................... max. 0.5×IΔN
Limit contact voltage ......................... 25 V or 50 V
The contact voltage is calculated to.. IΔN (general type) or to 2×IΔN (selective type)
Fault loop resistance at contact voltage is calculated as
N
C
LI
U
R
Δ
=.
100


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3102 BT - Eurotest XE at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3102 BT - Eurotest XE in the language / languages: English as an attachment in your email.

The manual is 1.18 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3102 BT - Eurotest XE

Metrel MI 3102 BT - Eurotest XE User Manual - Dutch - 33 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info