Do not start the device up until the clean components are completely dry.
7 If the compressor or outer section of the air tube is dirty, clean them
with a soft cloth moistened with water or mild detergent.
Mild Detergent
ON/OFF
The compressor is not water-proof.
8 Reassemble the nebulisation kit. Put it in the storage bag.
B. DISINFECTING THE NEBULISATION COMPONENTS
Always disinfect the drug reservoir, mouth piece and other nebulising components
after the last treatment of the day. To disinfect the components, follow one of the
methods described below:
B.1. Use a commercially available disinfectant. Follow the manufacturer’s instructions.
1 Immerse the components in the cleaning solution for the period indicated.
2 Remove the components and discard the solution.
3 Rinse the components thoroughly in hot tap water, remove
the surplus and allow to dry in the air in a clean place.
B. 2.
You can boil the components for 15 to 20 minutes. Carefully remove the components
after boiling, remove excess water and allow to dry in the air in a clean place.
C. HANDLING THE SPRAY CAP
The spray cap is an important component for spraying the drug.
Please take the following precautions when handling it.
Wash the spray cap after each use.
Do not use the brush, pin, etc. to clean the components.
If you disinfect the components in boiling water, make sure you boil them in large vol-
umes of water.
Do not boil the spray cap with items other than the accessories corresponding to the
nebulisation system.
Do not use the microwave oven, dishwasher drier or hairdryer to dry the device or its
components.
MAINTENANCE AND STORAGE
To keep your device in the best possible condition
and in order not to damage it, proceed as follows:
Do not leave the device unsupervised in the presence of young children or individuals
who cannot give their consent.
Do not subject the de vice or its co mponents to viole nt shocks such a s, for ex ample, drop ping
the device onto the ground.
Do not carry or withdraw the nebulisation kit when the drug reservoir contains a drug.
Do not store the device in a place t hat is subject to ex treme hot or cold temp eratures, hig h
humidity or in direct sunlight.
Do not bend or fold the air tube.
Dispose of the device and its optional used accessories or components according to the
local laws and regulations.
Never use benzene, thinner or ammable chemical product to clean the device.
Always carry the device in its storage bag and store it in the bag if you do not anticipate
using it for a long time.
CHANGING THE AIR FILTER
If the air lter has changed colour or been in regular use for over 60 days, replace it
with a brand new air lter.
1
2
3
1 Remove the compressor’s air lter cover.
2 Change the air lter.
3 Put the air lter cover back in situ.
Use only air lter s supplied w ith the d evice, whic h
are designed specically for this device. Do not
operate the unit without a lter.
The air lters can be inserted in any direction.
Check that the air lters are clean and f ree of du st
before inserting them.
Do not attempt to wash or dry the lter. If the
air lter is wet, replace it. Damp air lters may
cause obstructions.
To prevent the air lter cover becoming clogged,
wash it regularly. Do not let the air lter cover
boil. Remove the air lter cover, wash it and make
sure it is completely dry and attached correctly.
Please cont act your LANAFORM reseller to obtain additional air lters or if the components
supplied with the device are damaged or missing.
TROUBLE-SHOOTING
Check the following points if your unit malfunctions when running.
You can also consult the pages of these instructions for use for detailed instructions.
Symptoms Cause Solution
Nothing happens when
the mains switch is in
the ON position.
Have the batteries
been inserted into their
compartment or has the USB
cable been connected to the
compressor’s connector?
Check that the plug has
been inserted correctly
into the electric socket.
Nebulisation stops
automatically after 10 seconds
when the device is on.
The batteries are at.
Change the batteries or
connect the USB cable after
removing the batteries
from their compartment.
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Lanaform LA120420 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Lanaform LA120420 in the language / languages: English, German, Dutch, French, Italian, Spanish as an attachment in your email.

The manual is 15,36 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info