Getting Started
2-10 2001-900-01 Rev. K / August 2010
2. You can view the present selections of each
CONFIGURE DCV option by pressing the and
keys to move to the desired option, then pressing
ENTER. The present selection is shown blinking.
(Remember that additional information is available by
pressing the INFO key.)
This is the top level of the menu shown in Table 2-3.
Step 4
Configure for high accuracy
To make high accuracy readings of a voltage source, change
the configuration of the DC voltage measurement, as
follows:
1. Select SPEED from the CONFIGURE DCV menu, then
press ENTER. The multimeter displays the following
menu:
DCV MEASUREMENT SPEED
NORMAL FAST MEDIUM HIACCURACY
SET-SPEED-EXACTLY SET-BY-RSLN
2. Using the cursor keys, select the HIACCURACY option
(integration time of 10 power line cycles), then press
ENTER.
3. Select FILTER from the CONFIGURE DCV menu,
then press ENTER. The following menu is displayed:
DCV DIGITAL FILTER
AUTO AVERAGING ADVANCED
AVERAGING-MODE
4. Using the cursor keys, select the AVERAGING option,
then press ENTER. The instrument displays the
following:
AVG: 010 RDGS (1-100)
Use , , , , ENTER,EXIT, or INFO
5. This is a simple moving average filter of 10 readings,
which is the minimum recommended for high accuracy
DC voltage measurements. If it is sufficient, just press
ENTER or EXIT. If you want the instrument to average
more readings, use the cursor keys and RANGE keys to
increase the value (up to 100 readings). Press ENTER to
make the change.
This step does not enable the digital filter.
That is done by pressing the FILTER key
while outside the menu structure.
6. Select RESOLUTION from the CONFIGURE DCV
menu. The following menu is displayed:
SET DCV RESOLUTION
AUTO 3.5d 4.5d 5.5d 6.5d 7.5d
7. Using the cursor keys, select 7.5 digits of resolution,
then press ENTER.
8. Press EXIT to return to the normal display. The reading
should reflect the speed and resolution changes.
(Changes will affect only the DCV function.)
q~ÄäÉ=OJP
CONFIGURE DCV menu structure
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
Measurement speed (integration time) menu:
Select 1 PLC (power line cycle, 16.67msec for 60Hz, 20msec for 50Hz and 400Hz).
Select 0.01 PLC.
Select 0.1 PLC.
Select 10 PLC.
Set integration time in PLC (0.01-10).
Default to setting appropriate for resolution.
FILTER
AUTO
AV E R A G I NG
ADVANCED
AVERAGING-MODE
Digital filter menu:
Default to filter appropriate for integration time.
Program a simple average filter (1-100 readings).
Program a simple average filter (1-100 readings) with a noise tolerance
window (0-100% of range).
Select moving average or repeating average mode.
RESOLUTION
AUTO
3.5d, 4.5d, 5.5d, 6.5d, 7.5d
Display resolution menu:
Default to resolution appropriate for integration time.
Select a specific resolution.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2001 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2001 in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info