15.13 Testing in Accordance with EN 60601 (Feature KA01)
The following leakage and auxiliary currents can be measured in accor-
dance with the above mentioned standard in the operating mode, as well
as under normal and single-fault conditions:
Protective conductor measurement R
PE
,
test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
Earth leakage current I
PE
Housing leakage current I
HL
Patient leakage current I
PL
(with nominal voltage at application part)
Patient auxiliary current I
PA
Leakage current is converted to reference voltage (see limit values chapter
8 on page 18). Reference voltage must be adapted to the supply voltage
range.
Check connection parameters and start test.
To
Test Socket This is the default setting. For other connection types, please
refer to chapter 7 on page 17.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II). In all other
cases, or if it is not clear whether or not all exposed, conduc-
tive parts are connected to one another or to the protective
conductor, the safety class can be selected manually.
Type The DUT type can be selected from a list.
If the limit values of the 3rd issue are to be taken into account,
select a DUT type ending with ... 3rd under this parameter.
Cond. Various test conditions can be activated here, including
insulation resistance measurement.
App PrtSee below and on page 54.
ID-No. See parameters database in chapter 15.2 on page 37.
Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.
Configure Application Parts
Entry is made here indicating whether or not the application parts are to
be tested. Jacks A through K (for connection of cables or probes) can
also be assigned to groups (application parts) for mutual testing.
Selecting Preset Test Combinations
õ Use the cursor to select test combinations with groups of 1, 2, 5 or 10
application parts and acknowledge with the key.
The groups are assigned automatically after the application parts have
been selected.
Selecting User Defined Test Combinations
õ Select the respective application part from the JACK column with
the cursor and acknowledge with the key. You can set up a user
defined group for each application part which includes 1 to 10
application parts with the cursor keys in the GR column (GRoup).
Acknowledge your settings with the key.
As long as at least one group number has been entered, the application
parts test is preset in the “To Socket” window.
The application part type with the strictest limit value takes precedence
in the initial window. All groups are set to this type. Various types can be
assigned to the groups with “direct print” (adjustable option in the setup
menu).
If testing is not to be performed with application parts, the group assign-
ment must be cancelled with the “clear” function.
The TYPE is filled in automatically if the safety class has already been
entered to the “To Socket” window.
Flowchart, see next page.
56


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUTEST SIII at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUTEST SIII in the language / languages: English as an attachment in your email.

The manual is 0,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info