15.12 Testing in Accordance with IEC 62353/VDE 0751
The following measurements can be performed in accordance with the
above mentioned standard:
Protective conductor measurement R
PE
Test current: 200 mA DC,
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
Insulation measurement (can be additionally activated)
– R-INS LN-PE
(insulation resistance, LN to protective conductor)
– R-INS AP-PE (insulation resistance, app. part to protective conductor)
Equivalent device leakage current I
EDL
Equivalent patient leakage current I
EPL
Device leakage current (direct or residual current)
Contact current
Patient leakage current
(direct or mains at application part ) (only with Feature J01)
Leakage current is converted to reference voltage (see limit values chapter
8 on page 18). Reference voltage must be adapted to the supply voltage
range.
Check connection parameters and start test.
To
Socket This is the default setting. Refer to chapter 7 on page 17 for
other types of connection.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II).
In all other cases, or if it is not clear whether or not all
exposed, conductive parts are connected to one another or
to the protective conductor, the safety class can be selected
manually.
Type Select the device type (DUT) from the list.
If “old devices” is selected, limit values from DIN VDE 0701-
0702 are used.
AP-Type:... (BF): Application parts are recognized automatically, and can
be manually changed as well:
Select the App. prt. ... line with the
or key, acknowl-
edge with the
key and change with the or key.
02: The number of configured groups is displayed here.
After selecting the App. prt. ... line with the or key,
acknowledging twice with the key and then pressing the
or key, the “Configure application parts” menu appears
(see chapter 15.13 on page 56).
ID No. See parameters database in chapter 15.2 on page 37.
Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.
54


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUTEST SIII at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUTEST SIII in the language / languages: English as an attachment in your email.

The manual is 0,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info