52 GMC-I Messtechnik GmbH
15.11 Testing Devices in Accordance with EN 60335
The following tests can be performed in accordance with VDE 0700, part
500, (as part of DIN EN 50106:1998) in compliance with special rules for
routine testing of devices for which EN 60335-1 and EN 60967 apply:
Testing of the protective conductor connection by means of resistance
measurement R
PE
(permanent connection or with plug)
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
Testing for dielectric strength by applying DC high-voltage (1.5-fold value)
(Prerequisite: Feature F02 or SECUTEST SIII+H)
•Function test
The following tests can be performed in accordance with
EN 60335-1:1994:
Testing for dielectric strength by applying high-voltage
(Prerequisite: Feature F02 or SECUTEST SIII+H)
Testing for Equivalent Leakage Current
Additional testing options:
Insulation resistance measurement R
INS
Residual current
Check connection parameters and start test.
If desired, the test sequence can be adapted to the respective device
under test in the following menu in Setup... under Sequence....
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the
setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module
(accessory), or can be read in with a barcode scan-
ner (accessory).
If an incorrect entry is made: Only complete line can
be deleted, and deletion is only possible with the
key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.
52


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUTEST SIII at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUTEST SIII in the language / languages: English as an attachment in your email.

The manual is 0,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info