571678
15
Zoom out
Zoom in
Previous page
1/72
Next page
GMC-I Messtechnik GmbH 15
5 Classification of Devices Under Test
5.1 Safety Classes
Devices assigned to all of the following safety classes are equipped with
basic insulation, and provide for protection against electrical shock by
means of various additional precautions as well.
Safety Class I Devices
Exposed, conductive parts are connected to the protective conductor so
that they are not charged with voltage if the basic insulation should fail.
Safety Class II Devices
These devices are equipped with double insulation or reinforced insula-
tion.
Safety Class III Devices
and Devices with Internal Power Supply
These devices are powered with Safety Extra Low Voltage (SELV). Beyond
this, no voltages are generated which exceed SELV. These devices may
not be connected to the mains. They may only be connected to the test
instrument at jacks 1 through 3.
Note: The DUT may only be connected to jacks 1 to 3 at the test instru-
ment. It is only possible to perform a visual inspection, a measurement of
the insulation resistance or the supply voltage, see parameter “SC III U
V
on page 39.
Classification Parameter (in the Sequence ... menu)
The test instrument always performs testing in accordance with the strict-
est limit values of the respectively selected safety class. The test is failed if
this limit value is exceeded.
However, higher limit values are allowed for certain devices under test.
If the classification parameter has been activated (= x), the user is asked if
higher limit values are allowable for these devices. If the user responds
with “Yes”, the DUT is reevaluated and the test may be passed.
Devices with Internal Power Supply
Devices with internal power supply are tested like permanently connected
Safety Class II or III devices.
5.2 Application Parts (electrical medical devices)
Type B Application Parts (body)
Devices of this type are suitable for both internal and external patient
applications, except for use in direct proximity to the heart.
These devices provide for adequate protection against shock especially
as regards:
Reliable leakage current
Reliable protective conductor connection if utilized
Devices of the following safety classes are allowable:
I, II, III or devices with internal electrical power supply.
Type BF Application Parts (body float)
Same as type B, but with type F insulated application parts.
Type CF Application Parts (cardiac float)
Devices of this type are suitable for use directly at the heart. The applica-
tion part may not be grounded.
Devices of the following safety classes are allowable:
I, II or devices with internal electrical power supply.
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUTEST SIII at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUTEST SIII in the language / languages: English as an attachment in your email.

The manual is 0,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info