GMC-I Messtechnik GmbH 51
I
PE
PE Leakage Current
Alternative measurement:
DUT mains plug connected to the test socket
Sequence
Ð Select the test: I PE softkey.
Ð Connect the DUT to the test socket, or connect a clip-on am-
meter.
Ð Select measuring range, test voltage Ut with polarity, and mea-
surement duration.
Ð Select type of connection and type of current via measurement
type.
Ð Start the test: Press the ONSTART key.
Ð After each reconnection to the mains, and as soon as the first
test is started, a mains connection test is executed (see sec-
tion 4.1.1).
Ð Measured values are displayed.
Ð End the test: Press the ONSTART key.
Ð
Read the measured value and compare it with the table of
allowable limit values.
Maximum Allowable Limit Values for Leakage Current in mA
* For devices with heating power of greater than 3.5 kW
Key
NC normal condition
SFC single fault condition
Test Standard
Prot. conductor current Diff. current
NC SFC
VDE 0701-0702 3,5 mA / 1 mA/kW *
3,5 mA/ 1 mA/kW
*
IEC 62353
(VDE 0751-1)
51


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info