I
E
Equipment Leakage Current
Application
Measurement of device leakage current is required for:
Electrical medial devices in accordance with IEC 62353
(VDE 0751-1).
Definition
Devices leakage current is the sum of all leakage current from the
housing, exposed conductive parts and application parts to PE
(protective earth).
Parameters
Measuring range: 300 A, 3 mA, 30 mA or auto
(automatic switching amongst the ranges)
Test voltage Ut: 110 V ... 240 V
Test frequency ft: 400 Hz, 200 Hz, 60 Hz or 50 Hz
Measurement duration: 5 s, 10 s, 15 s, 30 s or continuous
Measurement type:
– Direct measuring method
1)
– Differential current measuring method
1)
– Equivalent leakage current measuring method
In order to select the correct type of connection:
Select the measurement type and check the associated wiring
diagram after accessing it with the help function.
1)
Measurement must be performed with mains polarity in both
directions. The largest value is docuimented.
52


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info