GMC-I Messtechnik GmbH 49
I
PE
PELeakage Current
Application
Measurement of protective conductor current may be performed
in the case of safety class I devices for which it cannot be assured
that all components exposed to line voltage are tested by means
of insulation resistance measurement, or if insulation resistance
measurement cannot be performed for other reasons.
Definition of Differential Current
Sum of instantaneous current values which flow via the L and N
conductors at the device mains connection. Differential current is
practically identical to fault current in the event of an error. Fault
current: Current which is caused by an insulation defect, and
which flows via the defective point.
Definition of Equivalent Measurement
Equivalent leakage current is current which flows through active
conductors of the device which are connected to each other (L/N)
to the protective conductor (SC1), or to exposed, conductive
parts (SC2).
Definition of Protective Conductor Current (direct measurement)
Current which flows through the protective conductor with
housings which are isolated from ground.
Differential Current Measuring Method
The device under test is operated with mains power. The sum of
the momentary values of all currents which flow through all active
conductors (L/N) at the mains side of the device connection is
measured. The measurements must be performed with mains
plug polarity in both directions.
Equivalent Measurement Method
A high-impedance power supply is connected between the
short-circuited mains terminals and all exposed metal parts of the
housing (which are connected to one another).
Current which flows over the insulation at the device under test is
measured.
Protective Conductor Current Measuring Method (direct measurement)
The device under test is operated with mains power. Current
which flows through the PE conductor to earth at the mains side
of the device connection is measured.
49


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info