15.3 Configuring Measuring Parameters
Depending upon the test regulation, various measuring parameters can
be configured for the test sequence (settings: x / – = function activated /
deactivated). All of the possible parameters for all of the regulations are
listed below. The Sequence... menu is accessed via the setup parameter in
the initial program window for the respective regulation.
Select the Sequence... menu and acknowledge.
Select a parameter, acknowledge, change and acknowledge the change.
Visual inspection This menu appears at the very beginning of the test
sequence.
Manual sequence Each test step must be acknowledged with the
key (see test sequence in chapter 8 on page 18
regarding test duration for automatic sequence).
Autostore After testing is completed, test data are automatically
stored to the test instrument(Feature KB01), or in the
(P)SI module (accessory).
Mains polarity
reversal L an N are reversed each time line voltage is applied
to the test socket.
Classification Questions regarding classification appear if limit
values are exceeded, see chapter 5 on page 15.
SC III U
V
Supply voltage is measured instead of insulation
resistance for active devices under test.
R-INS LN-PE Insulation resistance measurement is performed
between phase and neutral conductors, and the
protective conductor.
R-INS AP-PE Insulation resistance measurement is performed from
an application part to the protective conductor.
HV test A high-voltage test is performed
(prerequisite: Feature F02 or SECULIFE ST
HV)
HV test duration Duration of a high-voltage test (5 ... 60 s)
Auto (test) method The instrument recognizes whether or not the device
under test can be switched on:
Leakage or residual current, or insulation resistance
and earth leakage current are measured accordingly.
R-PE AC > 10 A
Protective conductor measurement with 200 mA DC
,
10 A AC
(feature G00 or
SECULIFE ST)
oder 25 A AC
(
feature
G01 or
SECULIFE ST HV)
R-PE with clip Protective conductor resistance can be determined
with the help of the WZ12C current clip
No I
HL
for SCI Contact current is not tested for SCI.
Mains wait A waiting period between switching mains power on
and starting the test can be entered here, e.g. in
order to suppress the display of recorded values
which occur during warm-up of the device under
test.
Patient auxiliary
current Measurement is performed with patient auxiliary
current
Adapter for socket Limit values are activated for permanently connected
devices. A device under test which is normally per-
manently installed can be connected to the test
socket via an adapter. No voltage may be applied to the
test socket when this test method is used.
First measured
values A menu for the entry of the first measured value
appears during the test sequence.
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info