•Perform function test if required:
The function test can be performed each time a safety test has been
successfully completed. The blinking signal lamp indicates that the
function test should be started. Beyond this, the function test can also
be started from the Function Test selector switch position. See chapter
14 on page 34 regarding performance of the function test.
•Display test results
(the worst measured values for the respective test sequence)
Save test results and print if required
15.2 Setting Up Test Sequences
The order of a test sequence cannot be changed!
However, it is possible to change limit values and measuring parameters,
and test steps can be added or skipped.
Principally, two test sequences are distinguished which are described on
the following page:
Sequence 1 (repair tests and periodic tests)
Sequence 2 (type tests)
Settings for the Test Sequence
Settings for the test sequence are not stored under a type designation,
but apply generally for the preset sequence of the respective test stan-
dard.
Type-related Settings (Type*)
(for tests per EN 60950, EN 61010 and EN 60335 only)
The type-related settings only include limit values. These limit values can
be stored for each type under an individual type designation. Measure-
ments the limit values of which have been deleted are skipped.
A type can be stored as default type. This type is indicated if the rotary
switch is set to Auto.
If no default type has been stored, the following appears for type in the
Auto rotary switch position: GENERAL and the general limit values of the se-
lected test standard apply.
Overview of adjustable measuring parameters (see chapter 15.3 for their meaning)
Adjustable measuring
parameters for test
sequences according
to standard
DIN VDE 0701
Part 1
DIN VDE 0701
Part 240
DIN VDE 0701-0702
DINEN60950
DINEN61010
DINEN60335
IEC 62353
IEC 601/EN 60601
can be stored
under
Measuring parameters of the initial program window
To test socket ••••••••
Adapter for socket ••••••• Test standard
(safety) class ••••••••
Ext. Cable WITH EL1 ————— Test standard
Combined Testing —————— Test standard
R-INS LN-PE —— ————— Test standard
AP-Type —————— •• Test standard
Limit values •••••••Test standard (Type)
Type •∗ •∗ •∗ Type
Measuring parameters in the Test Cond. menu (test conditions)
R-INS LN-PE ——————— Test standard
R-INS AP-PE ——————— Test standard
HV test ——————— Test standard
Patient Auxiliary Current
——————— Test standard
Measuring parameters in the Setup menu
Visual inspection •••••••• Test standard
Manual sequence •••••••• Test standard
Autostore •••••••• Test standard
Mains polarity reversal •••••••• Test standard
Classification •••••••• Test standard
SK III U
V
•••••••• Test standard
R-INS LN-PE •• •••• Test standard
HV test ——— ••• Test standard
HV test duration ——— ••• Test standard
Auto (test) method •••——— Test standard
R-PE AC > 10 A •••••• Test standard
R-PE with clip ——————— Test standard
No I HL for SC I ——————— Test standard
Mains wait •••••••• Test standard
First measured values
—————— Test standard
R-INS AP-PE —————— Test standard
See chapter 8 for device parameters in the test sequence
37


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info