36 GMC-I Messtechnik GmbH
15 Measurements in Accordance with National and
International Standards with Selector Switch in Auto Position
If measurements need to be performed in accordance with given stan-
dards which require specific tests, and if results need to be stored and
documented with a test report, an automatic test sequence is advisable
instead of individual measurements.
Note
+
for tests per EN 60950, EN 61010 and EN 60335
The DUT must be connected to the test socket for the automatic
test sequence! The DUT must comply with safety class I or II.
õ Connect the test instrument to mains power.
A mains connection test is initialized (see chapter 3.2 on page 12).
õ
Connect the DUT to the test socket at the test instrument (see
chapter 7 on page 17). The test instrument initializes
connection type recognition
.
õ Set the selector switch to the Auto position. If the DUT has been con-
nected to the test socket, safety class recognition is performed. Other-
wise, the safety class must be specified manually. In the initial program
window, move the cursor up with the
key into the third line and ac-
knowledge with the key. A safety class can now be selected with
the
and keys and acknowledged with the key.
õ In order to select the regulation for which testing is to be performed,
move the cursor up within the initial program window with the
key
into the top line and acknowledge with the key. The desired stan-
dard can now be selected with the
and keys and acknowledged
with the key. If you always want to test in accordance with a spe-
cific standard, this can be stored as the default standard in the setup
menu. Otherwise, the selected standard only remains active until the
test instrument is disconnected from mains power (factory default set-
ting: VDE 0701-0702).
õ The test sequence can be configured in the Setup menu, limit values
can be changed if necessary and database options can be configured.
õ The test sequence is started by selecting Start test and by acknowledg-
ing with the key (see “Test Sequences”).
Tests which have already been included in chapters 9 through 14 are not
described here again. The only exception is measurement of extension
cables.
15.1 Test Sequences
Test sequences for the various standards are always run in the same
order, assuming that the device under test has been properly connected
and the mains connection test has been passed.
The test sequence can be run step by step with manual activation of each
subsequent step if this function has been specified, or automatically. Step
by step manual operation can be selected in the initial program window in
the setup menu under “Sequence” ... “manual sequence”.
Safety class: SC I or SC II is automatically recognized by the test instru-
ment. If the safety class is not correctly recognized, you can change it
manually. This manual correction is not stored however.
Visual inspection: If “visual inspection” has been activated
in the initial program window under “Sequence”.
If a part is recognized as defective by the user, it must be identified as
such by selecting it with the cursor and acknowledging with the key.
Protective conductor resistance measurement
(for SC I devices under test only)
Note
+
This test cannot be skipped. Exception:
This measurement can be skipped by pressing
(when mes-
sage „Connect probe to protective conductor“ is displayed) if a
protective conductor connection is not possible.
Evaluation of protective conductor testing
High-voltage test for DIN VDE 0701 Part 1, Appendix E, EN 60950,
EN 61010, EN 60335, EN 60601, if selected in the setup menu.
Insulation resistance measurement
IEC 62353: only if selected in the setup menu of the
initial program window under “Sequence ...
EN 60601: only if selected in the initial program window
under „Test Conditions“
DIN VDE 0701-0702 only if R-INS is activated in the initial program
window
Evaluation of insulation test
Leakage current measurement (various single-fault conditions are run de-
pending on classification)
Evaluation of each individual leakage current measurement, see also
chapter 21.2
Evaluation of the overall test
36


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info