14 Function Test
The device under test can be subjected
to a function test with line voltage via the
integrated test socket.
In addition to testing with the selector switch in the function test position,
a function test can also be performed immediately after safety testing has
been passed in accordance with the selected standard (not possible for
safety class III devices).
Attention!
!
The function test may only be performed after the DUT has
successfully passed the safety test.
Note
+
Each time line voltage is applied to the test socket, phase con-
ductor L and neutral conductor N are reversed, assuming that
“polarity reversal = X” is selected in the I leakage switch setting.
Note
+
The function test is only possible if the device under test has
been connected to the test socket (21).
Measurements
The function test includes the following measurements:
–Voltage U
LN
between the L and N conductors
Residual current ΔI
(corresponds to fault current between L and N)
Load current I
a
Active power P
Apparent power AP (calculated)
Power factor PF (cos ϕ calculated, display > 10 W)
Electrical energy W
On-time t for U
LN
at socket (21)
The following values are also displayed in all test sequences in switch po-
sition AUTO after the function test has been completed:
Maximum residual current ΔI
max
Maximum load current I
Lmax
Maximum active power P
max
Power factor is calculated from active power and apparent power.
Power factor corresponds to cos ϕ for sinusoidal quantities (line voltage
and load current).
Attention!
!
Starting the Function Test
For reasons of safety, the device under test must be switched
off before the function test is started. This precaution prevents
inadvertent start-up of a device under test which may represent
a hazard during operation, e.g. a circular saw or a disc grinder.
Ending the Function Test
After completion of the function test, devices under test must be
turned off with their own switch – especially devices with
relatively high inductivity.
Function
Test
34


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info