18.3 Fuses
If a fuse has blown due to overloading, a corresponding message
error appears at the display panel. The instruments voltage mea-
suring ranges are nevertheless still functional.
Fuses– FUSE Message
These fuses are active during all measurements except for voltage
measurement.
Attention!
!
Disconnect the instrument from the measuring circuit be-
fore opening the battery compartment lid in order to re-
place the fuse (refer to page 3 for location)!
Checking the Fuses
If interruption of the test current circuit is detected either before or
during measurement, the "fuse" message appears at the LCD.
The message is cleared after pressing any key.
After the cause of error has been eliminated and the blown fuse
has been replaced, the measurement can be performed again
without an error message.
Attention!
!
Severe damage to the instrument may occur if incorrect
fuses are used.
Only original fuses from GMC-I Messtechnik GmbH may
be used (order no. 3-578-285-01 / SIBA 7012540.3.15
SI-EINSATZ FF 3.15A/600V 6.3X32).
Only original fuses assure required protection by means
of suitable blowing characteristics. Short-circuiting of
fuse terminals or the repair of fuses is prohibited, and is
life endangering!
The instrument may be damaged if fuses with incorrect
ampere ratings, breaking capacities or blowing charac-
teristics are used!
Replacing the Fuses
Open the battery compartment lid by loosening the two
screws.
Remove the blown fuse and insert a new one. A replacement
fuse is included in the battery compartment.
Insert the new fuse.
Replace the battery compartment lid and retighten the
screws.
18.4 Housing
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. In par-
ticular for the protective rubber surfaces, we recommend a moist,
lint-free microfiber cloth. Avoid the use of cleansers, abrasives or
solvents.
Returns and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German electrical and
electronic device law). This device is subject to the RoHS direc-
tive. We also make reference to the fact that the current status in
this regard can be accessed on the Internet at www.gossen-
metrawatt.com by entering the search term WEEE.
In accordance with WEEE 2012/19/EU and ElektroG, we
identify our electrical and electronic devices with the sym-
bol in accordance with DIN EN 50419 which is shown at
the right. Devices identified with this symbol may not be disposed
of with the trash. Please contact our service department regard-
ing the return of old devices (see address in section 20).
If the batteries or rechargeable batteries used in your instrument are
depleted, they must be disposed of properly in accordance with
valid national regulations.
Batteries may contain pollutants and heavy metals such as lead
(Pb), cadmium (Cd) and mercury (Hg).
The symbol at the right indicates that batteries must not
be disposed of with the trash, and must be brought to a
designated collection point.
Pb Cd Hg
53


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt Profitest Intro at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt Profitest Intro in the language / languages: English as an attachment in your email.

The manual is 1.69 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info