776259
51
Zoom out
Zoom in
Previous page
1/62
Next page
GMC-I Messtechnik GmbH 51
Reference Conditions
Line voltage 230 V ±0.1%
Line frequency 50 Hz ±0.1%
Meas. quantity frequency
45 Hz 65 Hz
Measured qty. waveform
Sine (deviation between effective and
rectified value 0.1%)
Line impedance angle cos ϕ=1
Supply voltage 12 V ±0.5 V
Ambient temperature +22 °C±3 K
Relative humidity 45% ±10%
Nominal Ranges of Use
Voltage UN120 V (108 ... 132 V)
230 V (196 ... 253 V)
400 V (340 ... 440 V)
Frequency fN16 Hz (15.4 ... 18 Hz)
50 Hz (49.5 ... 50.5 Hz)
60 Hz (59.4 ... 60.6 Hz)
200 Hz (190 ... 210 Hz)
400 Hz (380 ... 420 Hz)
Overall voltage range U
Y
65 ... 550 V
Overall frequency range 15.4 ... 420 Hz
Waveform Sinusoidal
Temperature range 0 °C ... + 40 °C
Supply voltage 8 ... 12 V
Line impedance angle Corresponds to cos ϕ= 1 ... 0.95
Power Supply
(Rechargeable) batteries 8 each AA 1.5 V
We recommend using the battery pack
(article number: Z502H).
Number of measurements (standard setup with illumination)
– For RINS 1 measurement – 25 s pause:
approx. 1100 measurements
– For RLO Auto polarity reversal / 1 Ω
(1 measuring cycle) – 25 s pause:
approx. 1000 measurements
Battery test Symbolic display of rechargeable bat-
tery voltage
Power management
Display illumination can be switched off.
The test instrument is switched off
automatically after the last key opera-
tion. The user can select the desired
on-time.
Safety shutdown If supply voltage is too low (U < 8.0 V),
the instrument is switched off, or can-
not be switched on.
Recharging socket Installed, optional rechargeable batter-
ies can be recharged directly by con-
necting a charger to the recharging
socket:
Z502R charger
Charging time Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
Overload Capacity
UL-PE, UL-N 600 V continuous
RCD, RE440 V continuous
ZL-PE, ZL-N 550 V (Limits the number of measure-
ments and pause duration. If overload
occurs, the instrument is switched off
by means of a thermostatic switch.)
RLO Electronic protection prevents switching
on if interference voltage is present.
Protection with
two fine-wire fuse FF 3.15 A 10 s,
Fuses blow at > 5 A
Electrical Safety
Protection class II per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal voltage 230/400 V (300/500 V)
Test voltage 3.7 kV 50 Hz
Measuring category CAT III 600 V or CAT IV 300 V
Pollution degree 2
Fuses
L and N terminals 1 cartridge fuse-link ea.
FF 3.15 A/600 V 6.3 x 32 mm
Electromagnetic Compatibility (EMC)
Product Standard EN 61326-1:2006
Ambient Conditions
Accuracy 0 to + 40° C
Operation -5 ... + 50°C
Storage -20 ... + 60°C
(without batteries)
Relative humidity
Max. 75%
(max. 85% during storage/transport)
no condensation allowed
Elevation Max. 2000 m
Calibration interval 1 year (recommended)
Mechanical Design
Display Multiple display with dot matrix,
128 x 128 pixels,
backlit (transflective),
dimensions: 65 x 65 mm
Dimensions W x L x D: 225 x 130 x 140 mm
Weight Approx. 1.5 kg with batteries
Protection Housing: IP 52, connections: IP 40
per EN 60529/DIN VDE 0470-1
Excerpt from Table on the Meaning of IP Codes
Data Interfaces
Type USB slave for connection to a PC
Type RS-232 for barcode and RFID readers
BAT
Interference emission Class
EN 55022 A
Interference immunity Test Value Fe at ure
EN 61000-4-2 Contact/atmos. –
4 kV/8 kV
EN 61000-4-3 10 V/m
EN 61000-4-4 Mains connection –
2 kV
EN 61000-4-5 Mains connection –
1 kV
EN 61000-4-6 Mains connection –
3 V
EN 61000-4-11 0.5 periods / 100%
IP XY
(1
st
digit X)
Protection Against
Foreign Object Ingress
IP XY
(2
nd
digit Y)
Protection Against Pene-
tration by Water
4 1.0 mm dia. 0 Not protected
5 Dust protected 2 Dripping (at 15° angle)
51


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt Profitest Intro at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt Profitest Intro in the language / languages: English as an attachment in your email.

The manual is 1.69 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info