776259
8
Zoom out
Zoom in
Previous page
1/62
Next page
8GMC-I Messtechnik GmbH
4 Initial Start-Up
4.1 Installing or Replacing the Battery Pack
Attention!
!
Before opening the battery compartment, disconnect the
instrument from the measuring circuit (mains) at all poles!
Note
See also section 18.2 on page 52 regarding the charging
procedure for the Compact Master Battery Pack (Z502H)
and concerning the Z502R charger.
If at all possible, use the Compact Master Battery Pack (Z502H) with
sealed cells which is available as an accessory. This ensures that the
complete set of rechargeable batteries is always replaced at the
same time and that all batteries are inserted with correct polarity,
in order to assure that they do not fail.
Only use commercially available rechargeable battery packs if they will
be externally recharged. The quality of these packs cannot be
checked and may result in overheating and thus deformation
under unfavorable conditions (when charging them in the instru-
ment).
Dispose of rechargeable battery packs or individual rechargeable
batteries in an environmentally sound fashion when their service
life has nearly expired (approx. 80% charging capacity).
Loosen the slotted screw for the rechargeable battery com-
partment lid on the back and remove the lid.
Remove the depleted rechargeable battery pack/holder.
Attention!
!
If the rechargeable battery holder is used:
Make sure that all of the batteries are inserted with cor-
rect polarity. If just one battery is inserted with reversed
polarity, it will not be recognized by the instrument and
may result in leakage from the batteries.
Individual rechargeable batteries may only be recharged
externally.
Insert the new rechargeable battery pack / loaded recharge-
able battery holder into the rechargeable battery compart-
ment.
The holder can only be inserted in its proper position.
Replace the lid and retighten the screw.
4.2 Switching the Instrument On/Off
The test instrument is switched on by pressing the ON/START key.
The menu which corresponds to the momentary selector switch
position is displayed.
The instrument can be switched off manually by simultaneously
pressing the MEM and HELP keys.
After the period of time selected in the SETUP menus has elapsed,
the instrument is switched off automatically (see “Device Set-
tings”, section 4.5).
4.3 (Rechargeable) Battery Test
If (rechargeable) battery voltage has fallen below the
allowable lower limit, the pictograph shown at the
right appears. “Low Batt!!!” is also displayed along with a
(rechargeable) battery symbol. The instrument does not function if
the batteries have been depleted excessively, and no display
appears.
4.4 Charging the Battery Pack in the Tester
Attention!
!
Use the Z502R charger in order to recharge the Compact
Master Battery Pack (Z502H) in the test instrument.
Make sure that the following conditions have been fulfilled be-
fore connecting the charger to the charging socket:
– The Compact Master Battery Pack (Z502H) has been
inserted, i.e. not a commercially available rechargeable
battery pack, individual batteries or non-rechargeable
batteries.
– The test instrument has been disconnected from the
measuring circuit at all poles.
– The instrument must remain off during charging.
Refer to section 18.2.1 with regard to charging a rechargeable
battery pack which has been inserted into the tester.
If the rechargeable batteries or battery pack have not been used
or recharged for a lengthy period of time (> 1 month), thus result-
ing in excessive depletion:
Observe the charging sequence (indicated by LEDs at the char-
ger) and initiate a second charging sequence if necessary (dis-
connect the charger from the mains and from the test instrument
to this end, and then reconnect it).
Please note that the system clock stops in this case and must be
set to the correct time after the instrument has been restarted.
BAT
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt Profitest Intro at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt Profitest Intro in the language / languages: English as an attachment in your email.

The manual is 1.69 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info