6GMC-I Messtechnik GmbH
1 Scope of Delivery
1Test instrument
1 Shoulder strap
1 Battery pack
1 KS-PROFiTEST INTRO (Z503L)
1 Factory calibration certificate
1 Condensed operating instructions
1 Supplementary sheet with safety information
1 Comprehensive operating instructions available on the Inter-
net for download at www.gossenmetrawatt.com
2 Application
This test instrument fulfills all requirements of applicable European
and national EC directives. We confirm this with the CE mark. The
relevant declaration of conformity can be obtained from
GMC-I Messtechnik GmbH.
The measuring and test instrument allows for quick and efficient
testing of protective measures in accordance with DIN VDE 0100-
600:2008
(Erection of low-voltage installations; tests – initial tests),
as well as
ÖVE-EN 1 (Austria), NIV/NIN SEV 1000 (Switzerland)
and other country-specific regulations.
The test instrument complies with IEC 61557/EN 61557/
VDE 0413 regulations:
Part 1: General requirements
Part 2: Insulation resistance
Part 3: Loop resistance
Part 4:
Resistance of earth connection and equipotential bonding
Part 5: Earth resistance
Part 6: Effectiveness of residual current devices (RCDs) in TT and
TN systems
Part 7: Phase sequence
Part 10:Electrical safety in low-voltage systems up to 1000 V AC
and 1500 V DC – Equipment for testing, measuring or
monitoring of protective measures
The test instrument is especially well suited for:
•Systems setup
Initial start-up
Periodic testing
Troubleshooting in electrical systems
All of the values required for approval reports (e.g. per ZVEH) can
be measured with this test instrument.
All acquired data can be archived, and measurement and test
reports can be printed out at a PC. This is of special significance
where product liability is concerned.
The applications range of the test instrument covers all alternating
and three-phase current systems with nominal voltages of 230 V /
400 V (300 V / 500 V) and nominal frequencies of 16⅔ / 50 / 60 /
200/ 400Hz.
The following can be measured an tested with the test instrument:
Voltage / frequency / phase sequence
Loop impedance / line impedance
Residual current devices (RCDs)
Earth resistance / earth loop resistance (relative to the mains)
Insulation resistance
Low-value resistance (potential equalization)
Voltage drop
Refer to section 19.3 regarding testing of electrical machines in
accordance with DIN EN 60204.
Refer to section 19.3 for periodic testing in accordance with
DGUV regulation 3 (formerly BGV A3).
2.1 Using Cable Sets and Test Probes
KS-PROFiTEST INTRO (Z503L)
Remote control with measurement key (Z550A), optional
accessory
Measurements per DIN EN 61010-031 may only be performed in
environments in accordance with measuring categories III and IV
with the safety cap attached to the test probe at the end of the
measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps
have to be removed by prying open the snap fastener with a
pointed object (e.g. the other test probe).
See also section 17.1, “Technical Data for Measurement Cables
and Adapters”, beginning on page 52.
2.2 Overview of Included Features
1The so-called live measurement is only advisable if there is no bias current within
the system. Only suitable for motor protection switches with small nominal current
values.
2Currently available languages: D, GB, I, F, E, P, NL, S, N, FIN, CZ, PL
PROFITEST INTRO (M520T)
Testing of residual current devices (RCDs)
UB measurement without tripping the RCD
Tripping time measurement
Measurement of tripping current IF
Selective, SRCDs, PRCDs, type G/R
AC/DC sensitive RCDs, types B and B+, EV, MI
Testing for N-PE reversal
Measurement of loop impedance ZL-PE / ZL-N
Fuse table for systems without RCDs
Without tripping the RCD, fuse table
With 15 mA test current1, without tripping the RCD
Earth resistance RE (mains operation)
Measurement of equipotential bonding RLO
Automatic polarity reversal
Insulation resistance RINS
Variable or rising test voltage (ramp)
Voltage UL-N / UL-PE / UN-PE / f
Special measurements
Phase sequence
Earth leakage resistance RE(INS)
Voltage drop (ΔU)
Features
Selectable user interface language 2
Memory (database for up to 50,000 objects)
RS 232 port for RFID/barcode reader
USB port for data transmission
ETC user PC software
Measuring category: CAT III 600 V / CAT IV 300 V
Factory calibration certificate
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt Profitest Intro at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt Profitest Intro in the language / languages: English as an attachment in your email.

The manual is 1.69 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info