GMC-I Messtechnik GmbH 75
Battery Test
is dis-
played
All
Rechargeable batteries must be recharged, or replaced towards the end
of their service life (U < 8 V).
PE test by means of finger contact at the contact surfaces on the test plug
LCD LEDs
is displayed
U
L
/R
L
FI/RCD
light up
red
XX
U
(single-phase
measurement)
Potential difference 50 V between finger contact and PE (earth contact)
Frequency f 50 Hz
is displayed
U
L
/R
L
FI/RCD
light up
red
XX
U
(single-phase
measurement)
If L is correctly contacted and PE is interrupted (Frequency f 50 Hz)
Error Messages — LCD Pictographs
XX
All measurements
with protective
conductor
Potential difference U
L
between finger contact and PE (earth contact)
(Frequency f 50 Hz)
Remedy: inspect PE connection
Note: only when appears: measurement can nevertheless be
started by pressing the start key again.
XX
I
N
/ I
F
Z
L-N
/ Z
L-PE
/ R
E
1) Voltage too high (U > 253 V) for RCD test with direct current
2) U always U > 550 V with 500 mA
3) U > 440 V for I
N
/ I
F
4) U > 253 V for I
N
/ I
F
with 500 mA
5) U > 253 V for measurement with probe
XX I
N
RCD is tripped too early, or is defective.
Remedy: test circuit for bias current
XX Z
L-PE
RCD is tripped too early, or is defective.
Remedy: Test with “DC + positive half-wave”.
XX I
N
/ I
F
RCD tripped during contact voltage measurement.
Remedy: Check selected nominal test current.
EXTRA PRCD
PRCD has tripped.
Cause: poor contact or defective PRCD
X X All except for U
Externally accessible fuse is blown.
The voltage ranges remain functional even if fuses have blown.
Special case, R
LO
: Interference voltage during measurement may result in a
blown fuse.
Remedy: Replace fuse (replacement fuses in battery compartment).
Observe notes regarding fuse replacement in section 20.3!
XX
I
N
/ I
F
Z
L-N
/ Z
L-PE
/ R
E
Frequency out of permissible range
Remedy: inspect mains connection
Status
Test
plug
Meas.
Adapter
Position of the
Function Switch
Function / Meaning
PE
PE
75


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt PROFITEST MXTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt PROFITEST MXTRA in the language / languages: English as an attachment in your email.

The manual is 2,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt PROFITEST MXTRA

Gossen Metrawatt PROFITEST MXTRA User Manual - Dutch - 92 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info