4 Measuring Direct and Alternating Voltage
Direct voltages and sinusoidal alternating voltages with frequencies
ranging from 40 to 200 Hz can be measured with this instrument. Needle
deflection is always positive for direct voltage measurements regardless of
polarity at the terminal connections. Alternating voltage is displayed as an
RMS value.
Set the function selector switch to the “1000 V” position.
Make certain that the needle points to the “0” mark at the V scale in
the idle state. Readjust the needle if necessary with the adjusting
screw for mechanical zero adjustment.
Rotary switch position is irrelevant for voltage measurement.
Probe the measuring point with both test probes.
Read the measurement value from the V scale.
Notes
Allowable overload capacity for the voltage measuring range is
1200 V.
Input impedance for the voltage measuring range is 0.9 MΩ.
Devices under test with capacitive charges are automatically
discharged by the test instrument. The corresponding drop in voltage
can be observed at the display.
5 Measuring Insulation Resistance
Before performing measurement, make certain that the device under test
is voltage-free (see chapter 4).
5.1 Measurement Procedure
First set the function selector switch to the R1 position.
Briefly short circuit the test probes and press the measurement key to
check for full-scale deflection for R1 = 0 MΩ (zero adjustment).
The needle must point to the above indicated full-scale deflection
mark in the idle state. Readjust the needle if necessary with the rotary
knob for electrical zero adjustment.
Depending upon the nominal voltage of the device under test, select a
test voltage of either 50 V, 100 V, 250 V, 500 V or 1000 V with the
rotary switch.
Note
Full-scale deflection must be tested in the following cases, and must be
readjusted if required in order to attain a measuring accuracy of 1.5%:
Each time a different nominal voltage is selected
For resistance measurements of long duration
Probe the measuring point with both test probes.
Press the measurement key and read the measurement value from
the appropriate scale.
If needle deflection is insufficient, switch to the next highest
measuring range, i.e. R2 or R3, with the function selector switch.
Multiply the displayed value by the conversion factor which appears
opposite the selected nominal voltage:
Attention!
!
Do not touch the conductive ends of the two test probes after the
instrument has been activated for the measurement of insulation
resistance.
If measurement is being performed at a purely ohmic device
under test, current may flow over your body, which, although it
does not reach life endangering levels, causes a clearly
discernible shock.
When measurements are made at capacitive devices under test,
e.g. cables, the DUT may be charged with voltages of up to
approximately 1200 V, depending upon selected nominal voltage.
Contact with the device under test after measurement is, in this
case, life endangering!
For this reason, the DUT must be discharged in a controlled
fashion as described in chapter 4 on page 5.
Nominal Voltage: 50 V 100 V 250 V 500 V 1000 V
Conversion
Factor:
x 0.1 x 0.2 x 0.5 x 1 x 2
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO 1000A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO 1000A in the language / languages: English as an attachment in your email.

The manual is 0,26 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info