6 GMC-I Messtechnik GmbH
If the symbol is included in the description of the surge pro-
tection device, the manufacturer’s instructions must be observed
for the respective device type.
The actual measurement is performed as follows:
Select measurement of surge protection devices with the
R
ISO/USPD key, and select the desired sub-function with the
FUNC key (see also description above). Example:
Figure 3.11: Surge Protection Device Measurement Menu,
DC Function
Connect the test probes to the surge protection device in ac-
cordance with the manufacturer’s instructions.
Figure 3.12: Connection Example
Note
If a voltage of greater than about 10 V is present at the measured
surge protection device, its value is displayed and the START key is
disabled (see section 3.3.1 on page 4). Eliminate the source of
interference voltage before restarting the measurement.
Trigger measurement by pressing the START key. Release the
START key as soon as measurement is started. Rising current
at the surge protection device is displayed by means of a bar
graph. At the same time, the measuring voltage value appears
at the display in the U
IN field. Measurement is ended automat-
ically.
Read voltage measured at the milliampere point.
Note
The device under test may not be disconnected from the
test instrument as long as the “!” warning symbol is lit up.
After insulation measurement has been completed, any remaining
residual voltage is displayed as UIN, which may result from cable
capacitance. Contact to the device under test must be main-
tained for as long as the capacitive DUT is being discharged via
the test instrument’s internal resistor. The falling voltage value can
be observed directly at the UIN display. Do not disconnect the
DUT until voltage U
IN is less than 25 V!
Comments
Before measuring surge protection devices, disconnect them
from the installation.
It’s advisable to study the circuit before performing measure-
ment. Surge protection devices are currently equipped with
integrated interference suppression filters and the like, which
may influence measurement results.
3.4 Further Device Functions
Select Language, Query Firmware Version
Before selecting a language or querying the firmware version, dis-
connect both test probes from the device under test / measuring
circuit and switch the test instrument off.
Press and hold the R
ISO/USPD key while switching the test in-
strument on.
The firmware version and other service information appears at the
display, as well as the language selection menu.
Select the desired language with the corresponding key
(EN = English, CZ = Czech).
Figure 3.15: Language Selection Menu
The instrument is switched back to normal operation after a lan-
guage has been selected.
Measuring Point Illumination with White LED
The LED can be switched on and off by briefly pressing the START
key.
No voltage may be applied to the test probes.
!
Figure 3.13: Sample Results
for the USPD
Measurement
(discharging is
active: residual
voltage = 144 V)
Figure 3.14: Further Sample
Results for the
U
SPD Measure-
ment (DUT is fully
discharged: resid-
ual voltage = 0 V)
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRALINE ISO-CHECK at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRALINE ISO-CHECK in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRALINE ISO-CHECK

Gossen Metrawatt METRALINE ISO-CHECK User Manual - Dutch - 8 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info