4 GMC-I Messtechnik GmbH
3 Performing Measurements
3.1 Switching the Instrument On and Off, Energy-Saving
Mode, Automatic Shutdown
The instrument is switched on by pressing and holding the START
key. Briefly press the START key twice in order to switch the instru-
ment off, during which no voltage may be applied to the test
probes! The instrument is switched to the standby mode after
several seconds (reduced brightness), if none of the keys has
been pressed and no voltage is applied to the test probes. The
instrument is switched out of the standby mode (i.e. back to full
brightness) by pressing any key or applying voltage to the test
probes. The instrument is shutdown automatically if it has
remained inactive for about 1 minute, i.e. if no keys have been
pressed and no voltage has been applied to the test probes dur-
ing this time.
3.2 Instructions and Principles with Validity for
All Measurements
The desired functions or parameters are selected with the
R
ISO/USPD, FUNC, UTEST and UTEST keys. Measurement is
triggered by pressing the START key. All selected functions or
parameters remain valid until they are changed.
If (interference) voltage of greater than 10 V is applied to the
test probes, this is indicated at the display in the UIN field
along with the measured value, as well as the “!” symbol. If
this is the case, the START key is disabled.
Bild 3.1 (Interference) Voltage Display
If battery voltage is too low (only the red field is lit up in the
battery symbol), measurement cannot be started. After press-
ing the START key, the depleted battery symbol appears for
about 1 second (see figure below). Replace the batteries as
described in section 5.1 on page 7.
When measuring high insulation resistance, the measurement
cables should be laid unobstructed in the room, or placed
upon a material with good insulating properties.
Contact the device under test with the test probe before trig-
gering measurement with the START key, thus causing any
possible interference voltage to appear at the display.
Do not remove the test probes prematurely during the course
of the measurement, because this may lead to distorted mea-
surement results.
After insulation measurement has been completed, any remaining
residual voltage is displayed as U
IN , which may result from cable
capacitance. Contact to the device under test must be main-
tained for as long as a capacitive DUT is being discharged via the
test instrument’s internal resistor. The falling voltage value can be
observed directly at the U
IN display. Do not disconnect the DUT
until voltage U
IN is less than 25 V!
3.3 Measuring Functions
Attention!
Each time before measuring insulation resistance or surge
protection, make sure that the device under test is voltage-
free!
Do not touch the device under test or the test probes during
measurement, and for a short time thereafter as long the DUT
is being discharged.
The DUT may be charged with as much as 1000 V during
measurement, for which reason contact to the DUT must be
maintained after measurement has been completed until it is
fully discharged! Residual voltage appears at the display along
with the “!” warning symbol.
Do not remove the test probes until voltage has dropped to a
safe level and the “!” warning symbol has been cleared from
the display.
All consuming devices must be disconnected and the
switches must be turned on when measuring insulation resis-
tance between conductors.
3.3.1 Voltage Measurement
Connect the test instrument to the device under test.
Figure 3.4: Connection Example for Voltage Measurement
If a voltage of greater than about 10 V is present in the mea-
sured electrical circuit, its value is displayed in the U
IN field and
the “~” symbol appears for AC or polarity is displayed for DC.
The “+” symbol lights up if the positive pole is connected to
the fixed test probe, and “–” appears if the negative pole is
connected to the fixed test probe. “!” appears at the same
time as a warning. The START key is disabled!
Figure 3.2: Low Battery
Voltage
Figure 3.3: Low Battery
Voltage: Display
After Pressing the
START Key
Figure 3.5: Display of (Inter-
ference) Voltage
(Riso function)
Figure 3.6: Voltage
Measurement
(U
SPD function)
!
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRALINE ISO-CHECK at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRALINE ISO-CHECK in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRALINE ISO-CHECK

Gossen Metrawatt METRALINE ISO-CHECK User Manual - Dutch - 8 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info