590625
40
Zoom out
Zoom in
Previous page
1/74
Next page
40 GMC-I Messtechnik GmbH
V/Hz, Ω, Temperature, and A/Hz Measurements
5.5.2 Measurement with Resistance Sensors
Ð Set the rotary switch to “Temp
TC
” or “Temp
RTD
”.
The last selected temperature measurement or the last selected
temperature sensor, i.e. type K or Pt100/Pt1000, remains in
memory and is accordingly displayed. Press the FUNC | ENTER key
in order to change to the other measuring function.
The sensor type, i.e. Pt100 or Pt1000, is detected automatically
and displayed.
There are two different ways to compensate for cable resistance:
Automatic Compensation
Ð Acknowledge by pressing the ZERO | ESC key.
“Short leads” appears at the display.
If you prefer to enter cable resistance directly, you can skip the
following entry prompt.
Ð Short circuit the measuring instrument’s connector cables.
000.00” appears at the display. After pressing the FUNC | ENTER
key, automatic compensation of cable resistance is activated
for all subsequent measurements. The short-circuit can now
be eliminated, and the device is ready for use.
Entering Cable Resistance
Ð Press the ZERO | ESC key once again in the automatic compen-
sation menu.
Ð Enter the known resistance of the connector cables with the
scroll keys:
Select the digit to be changes with the
keys, and change
the respectively selected digit with the
 keys. The default
value is 0.43 Ω (Z3409). Values can be selected within a range
of 0 to 50 Ω.
Ð
Upon pressing the
FUNC | ENTER
key, the selected value is acti-
vated and the display is returned to the measuring function.
Cable resistance remains in memory even after the instrument
has been switched off.
COM
V
Temp
Ω
Temp
TC
Temp
RTD
TC
RTD
00
.
00
Ω
short
leads
Entering Cable Resistance
Automatic
RTD
Compensation
Measuring Ranges
RTD Pt 100 200,0 +850,0 °C
RTD Pt 1000 150,0 +850,0 °C
FUNC
ENTER
FUNC
ENTER
ZERO
ESC
ZERO
ESC
FUNC
ENTER
00
.
50
Ω
0023. .3
°C
TC
1
.
22.4
0256
.
0
°C
Pt1000
rleadsmeas
rleadsset
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT ULTRA BT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT ULTRA BT in the language / languages: English as an attachment in your email.

The manual is 0,9 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info